Olympus Suction Pump Suction Cleaning Adapter MAJ 222 User Manual

To use this chart as a wallchart, cut  
off the top of the chart as illustrated.  
To insert this chart into a binder, punch out  
the holes and fold the chart as illustrated.  
BRONCHOSCOPE DISINFECTION GUIDE  
EXERABF-160 Series and EVISBF-240 Series Bronchovideoscopes  
Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information  
WARNING: This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the  
detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased.  
HIGH-LEVEL DISINFECTION  
2
1
3
4
Suction Cleaning Adapter  
Fill basin with disinfectant. Attach a  
suction cleaning adapter (MAJ-222)  
to the previously cleaned broncho-  
scope. Immerse completely in  
disinfectant solution.  
Attach a syringe to the adapter.  
While immersed, detach the adapter  
Soak all equipment according to  
the time and temperature  
recommended by the disinfectant  
manufacturer. Use a timer to  
measure the exact contact time.  
Pull the plunger of the syringe to  
fill the instrument/suction channels  
and channel cleaning adapter  
with disinfectant solution.  
from the bronchoscope. Use a clean,  
lint-free cloth to wipe away any bub-  
bles clinging to external surfaces.  
RINSING  
To Suction Pump  
Suction Cleaning  
Adapter (MAJ-222)  
5
6
7
8
To Suction Pump  
Alcohol  
Connect the suction cleaning  
adapter and syringe. Inject air  
to remove disinfectant from the  
instrument/suction channels,  
then remove the bronchoscope  
from the disinfectant.  
Disconnect the suction cleaning  
adapter and immerse the  
bronchoscope and adapter in  
sterile water. Using a sterile,  
lint-free cloth, thoroughly rinse  
and wipe all surfaces.  
Connect suction cleaning adapter  
to bronchoscope and to a tube  
from a suction pump. Aspirate  
water through channels for 30  
seconds. Remove from water and  
aspirate air through channels for  
60 seconds.  
Immerse the bronchoscope's distal  
end in 70% alcohol and aspirate  
alcohol for 5 seconds. Remove the  
distal end from the alcohol and  
aspirate air for 20 seconds.  
DRYING  
TRANSPORTATION  
9
10  
Turn the suction pump off. Hold the control  
section with the openings pointing down and  
disconnect the suction cleaning adapter.  
Dry the bronchoscope and adapter with an  
alcohol-moistened cloth. Make sure that the  
equipment is completely dry prior to storage.  
Hold all parts of the endoscope when carrying. Do  
not let the ends of the bronchoscope swing freely.  
NOMENCLATURE  
IMPORTANT  
• Wear appropriate personal protective  
equipment during reprocessing.  
Single-Use Suction Valve (MAJ-209)  
Suction Cylinder  
Light Guide  
Universal Cord  
• Autoclave all autoclavable accessories.  
• Store the bronchoscope hanging vertically  
with all valves and caps removed.  
Remote Switches  
• Use a compatible disinfectant and test  
it for proper concentration as recom-  
mended by the manufacturer.  
Electrical Connector  
Venting Connector  
Single-Use Biopsy Valve (MAJ-210)  
Instrument/Suction Channel Port  
• Do not use the carrying case to store  
"patient-ready" endoscopes. Any broncho-  
scope placed in the carrying case must be  
reprocessed prior to patient use.  
Water-Resistant Cap  
(MH-553)  
Insertion Tube  
Control Section  
Boot  
Bending Section  
Clean the bronchoscope  
Distal Tip  
UP/DOWN Angulation Control Lever  
meticulously prior to disinfection.  

Philips Car Video System 192EL2 User Manual
Olympus EVOLT E 420 User Manual
Olympus E P3 User Manual
Nikon 2177 User Manual
Lochinvar SOLUTION CB 135 User Manual
LG Electronics E224TV User Manual
KitchenAid KFPW760QBW1 User Manual
Kicker ZX SERIES ZX25001 User Manual
Kenwood Car Satellite Radio System TK 940 User Manual
JVC KD SH909 User Manual