Polaroid CCD Camera User Manual

Operating Manual  
CCD Camera Models  
ST-7E, ST-8E, ST-9E, ST-10E  
and ST-1001E  
Santa Barbara Instrument Group  
147A Castilian Drive  
Santa Barbara, CA 93117  
Phone (805) 571-7244 · Fax (805) 571-1147  
Web:<www.sbig.com> · Email:<[email protected]>  
Table of Contents  
1.  
1.1.  
1.2.  
Introduction.................................................................................................................1  
Road Map of the Documentation ...............................................................................1  
Quick Tour....................................................................................................................1  
1.2.1. CCDOPS Software ....................................................................................2  
1.2.2. CCD Camera .............................................................................................3  
2.  
2.1.  
2.2.  
Introduction to CCD Cameras...................................................................................5  
Cameras in General .....................................................................................................5  
How CCD Detectors Work .........................................................................................5  
2.2.1. Full Frame and Frame Transfer CCDs....................................................6  
Camera Hardware Architecture .................................................................................6  
CCD Special Requirements.........................................................................................8  
2.4.1. Cooling.......................................................................................................8  
2.4.2. Double Correlated Sampling Readout ....................................................9  
2.4.3. Dark Frames ..............................................................................................9  
2.4.4. Flat Field Images.....................................................................................10  
2.4.5. Pixels vs. Film Grains .............................................................................10  
2.4.6. Guiding ....................................................................................................11  
Electronic Imaging.....................................................................................................12  
Black and White vs. Color.........................................................................................13  
2.3.  
2.4.  
2.5.  
2.6.  
3.  
At the Telescope with a CCD Camera....................................................................15  
Step by Step with a CCD Camera ............................................................................15  
Attaching the Camera to the Telescope ...................................................................15  
Establishing a Communications Link ......................................................................16  
Focusing the CCD Camera .......................................................................................16  
Finding and Centering the Object ............................................................................18  
Taking an Image ........................................................................................................18  
Displaying the Image ................................................................................................18  
Processing the Image.................................................................................................19  
Advanced Capabilities ..............................................................................................19  
3.9.1. Crosshairs Mode (Photometry and Astrometry) .................................19  
3.9.2. Sub-Frame Readout in Focus .................................................................19  
3.9.3. Track and Accumulate............................................................................20  
3.9.4. Autoguiding and Self Guiding ..............................................................20  
3.9.5. Auto Grab ................................................................................................21  
3.9.6. Color Imaging .........................................................................................21  
3.1.  
3.2.  
3.3.  
3.4.  
3.5.  
3.6.  
3.7.  
3.8.  
3.9.  
4.  
Camera Hardware .....................................................................................................23  
System Components..................................................................................................23  
Connecting the Power ...............................................................................................23  
Connecting to the Computer ....................................................................................23  
Connecting the Relay Port to the Telescope............................................................23  
4.4.1 Using Mechanical Relays ...........................................................................24  
Modular Family of CCD Cameras ...........................................................................26  
4.1.  
4.2.  
4.3.  
4.4.  
4.5.  
i
4.6 Connecting the older model CFW-6 filter wheel to the Camera ..................................30  
4.7 Battery Operation .............................................................................................................31  
4.8 ST-1001E Differences........................................................................................................31  
5.  
Advanced Imaging Techniques ..............................................................................33  
Lunar and Planetary Imaging...................................................................................33  
Deep Sky Imaging .....................................................................................................33  
Terrestrial Imaging ....................................................................................................33  
Taking a Good Flat Field...........................................................................................33  
Building a Library of Dark Frames ..........................................................................34  
Changing the Camera Resolution ............................................................................34  
Flat Fielding Track and Accumulate Images...........................................................35  
Tracking Functions ....................................................................................................36  
5.1.  
5.2.  
5.3.  
5.4.  
5.5.  
5.6.  
5.7.  
5.8.  
6.  
Accessories for your CCD Camera..........................................................................39  
Cooling Booster..........................................................................................................39  
Tri-color Imaging .......................................................................................................40  
Camera Lens Adapters and Eyepiece Projection ....................................................40  
Focal Reducers ...........................................................................................................40  
AO-7 and Lucy-Richardson Software ......................................................................41  
SGS - Self-Guided Spectrograph...............................................................................41  
Third Party Products and Services...........................................................................41  
6.7.1. Windows Software..................................................................................41  
6.7.2. Image Processing Software ....................................................................41  
6.7.3. Getting Hardcopy ...................................................................................41  
SBIG Technical Support.............................................................................................42  
6.1.  
6.2.  
6.3.  
6.4.  
6.5.  
6.6.  
6.7.  
6.8.  
7.  
Common Problems ...................................................................................................43  
Glossary......................................................................................................................45  
Appendix A - Connector ad Cables ........................................................................49  
8.  
A.  
A.1. Appendix A - Connector Pinouts.............................................................................49  
A.2. SBIG Tracking Interface Cable (TIC-78)...................................................................49  
B.  
B.1.  
B.2.  
Appendix C - Maintenance......................................................................................51  
Cleaning the CCD and the Window ........................................................................51  
Regenerating the Desiccant.......................................................................................51  
C.  
Appendix C - Capturing a Good Flat Field............................................................53  
C.1.  
Technique 53  
ii  
Section 1 - Introduction  
1.  
Introduction  
Congratulations and thank you for buying one of Santa Barbara Instrument Group's CCD  
cameras. The model ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E are SBIG's fourth generation  
CCD cameras and represent the state of the art in CCD camera systems with their low noise  
and advanced capabilities, including Kodak's new Blue Enhanced E series of CCDs. We feel  
that these cameras will expand your astronomy experience by being able to easily take images  
like the ones you've seen in books and magazines, of structure never seen through the  
eyepiece. SBIG CCD cameras offer convenience, high sensitivity, and advanced image  
processing techniques that film just can't match. While CCDs will probably never replace film  
in its large format, CCDs allow a wide range of scientific measurements and have established a  
whole new field of amateur astronomy that is growing by leaps and bounds.  
The ST-7E, ST-8E, ST-9E and ST-10E cameras include an exciting new feature:  
self-guiding (US Patent 5,525,793). These cameras have two CCDs inside; one for guiding and  
a large one for imaging. The low noise of the read out electronics virtually guarantees that a  
usable guide star will be within the field of the guiding CCD for telescopes with F/numbers  
F/6.3 or faster. The relay output plugs directly into most recent commercial telescope drives.  
As a result, you can take hour long guided exposures with ease, with no differential deflection  
of guide scope relative to main telescope, and no radial guider setup hassles, all from the  
computer keyboard. This capability, coupled with the phenomenal sensitivity of the CCD, will  
allow the user to acquire observatory class images of deep sky images with modest apertures!  
The technology also makes image stabilization possible through our AO-7, or self-guided  
spectroscopy with our SGS. Due to the large size of the imaging CCD, The ST-10E camera does  
not have a second CCD in the head for self-guiding. For this camera we recommend a separate  
autoguider like the STV.  
1.1. Road Map of the Documentation  
This manual describes the ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E CCD Camera Systems  
from Santa Barbara Instrument Group. For users new to the field of CCD Astronomy, Sections  
2, 3 and 4 offer introductory material about CCD Cameras and their applications in  
Astronomy. Users who are familiar with CCD cameras may wish to skip section 2 and browse  
through sections 3 and 4, reading any new material.  
Thoroughly experienced SBIG customers may wish to jump right to the separate  
Software Manual, which gives detailed and specific information about the SBIG software.  
Sections 5 and 6 offer hints and information about advanced imaging techniques and  
accessories for CCD imaging that you may wish to read after your initial telescope use of the  
CCD camera. Finally, section 7 may be helpful if you experience problems with your camera,  
and the Appendices provide a wealth of technical information about these systems.  
1.2. Quick Tour  
This section is a quick guided tour of the CCD Camera System you have just purchased. If  
you're like most people you want to get started right away and dispense with the manual. Use  
this section as a guide for learning about your new system.  
Page 1  
Section 1 - Introduction  
1.2.1. CCDOPS Software  
Follow the instructions below to run the CCDOPS software and display and process sample  
images included on the distribution diskette.  
Install the software onto your hard disk. For Windows this involves running  
the Setup.exe file on the first diskette. For Macintosh or DOS this involves  
copying the contents of the floppy disk to a folder or directory on your hard  
disk.  
CCDOPS for Windows or Macintosh  
Double-click on the CCDOPS icon to launch the program.  
Use the Open command in the File menu to load one of the sample images.  
A window showing the exposure time, etc. will appear. Click in it to make it  
disappear. The image will show up in its own window.  
Try using the crosshairs. Use the Crosshairs command in the Display menu.  
Use the mouse to move the crosshair around in the image and see the pixel  
values.  
Close the crosshairs and try inverting the image. Click the Invert item in the  
Contrast window.  
Try the photo display mode. Use the Photo Mode command in the Display  
menu. Click the mouse to return to the menus.  
Load up the other sample images and display them using the photo display  
mode. You have to close any existing image first.  
If you find that the display is too dark or bright, try setting Auto Contrast in  
the Contrast window or adjust the background and range parameters to  
achieve the best display. You may have to hit the Apply button in the  
Contrast window to see changes in the Background and Range  
CCDOPS for DOS  
You should make the CCDOPS floppy or directory the active directory, then  
give DOS the CCDOPS command.  
Use the Open command in the File menu to load one of the sample images.  
Use the Image command in the Display menu to display the image.  
Try using the crosshairs. Hit the 'X' key. Use the mouse or arrow keys to  
move the crosshair around in the image and see the pixel values.  
Quit the crosshairs and try inverting the image. Hit the Esc key to quit the  
Crosshairs mode, then hit the 'N' key.  
Try the photo display mode. Exit the display mode by hitting the Esc key  
and then use the Display Image command again, this time selecting the  
Photo Display mode instead of the Analysis mode.  
Page 2  
Section 1 - Introduction  
Load up the other sample images and display them using the photo display  
mode.  
If you find that the display is too dark or bright, try setting Auto Contrast in  
the display menu or adjust the background and range parameters to achieve  
the best display. Usually your monitor brightness and contrast want to be  
set fairly high.  
Note: Full daylight at F/22 will saturate these cameras with the shortest exposure.  
With a camera lens start out in dim room light. For full sunlight you will need  
a neutral density filter.  
1.2.2. CCD Camera  
Unfortunately there really aren't many shortcuts you can take when using the CCD camera to  
capture images. The instructions below refer you to various sections of the manual.  
Insert the CCD Camera into the telescope and focus on a star (refer to  
Sections 3.2 and 3.3).  
Find some relatively bright object like M51, the Ring Nebula (M57) or the  
Dumbbell Nebula (M27) (refer to section 3.5).  
Take a 1 minute exposure using the Grab command with the Dark frame  
option set to Also (refer to Section 3.6).  
Display the image (refer to Section 3.7).  
Process the image (refer to Section 3.8).  
If you happen to have purchased a camera lens adapter for your CCD Camera you can use that  
to take images in the daytime. Additionally you could make a small pinhole aperture out of a  
piece of aluminum foil after wrapping it around the camera's nosepiece.  
Shut down the F stop all the way to F/16 or F/22.  
Set the focus based upon the object and the markings on the lens.  
Take a 1 second exposure with the Grab command.  
Display the image (refer to Section 3.7).  
Process the image (refer to Section 3.8).  
Page 3  
Section 2 - Introduction to CCD Cameras  
2.  
Introduction to CCD Cameras  
This section introduces new users to CCD (Charge Coupled Device) cameras and their  
capabilities and to the field of CCD Astronomy and Electronic Imaging.  
2.1. Cameras in General  
The CCD is very good at the most difficult astronomical imaging problem: imaging small, faint  
objects. For such scenes long film exposures are typically required. The CCD based system  
has several advantages over film: greater speed, quantitative accuracy, ability to increase  
contrast and subtract sky background with a few keystrokes, the ability to co-add multiple  
images without tedious dark room operations, wider spectral range, and instant examination of  
the images at the telescope for quality. Film has the advantages of a much larger format, color,  
and independence of the wall plug (the SBIG family of cameras can be battery operated in  
conjunction with a laptop computer, though, using a power inverter). After some use you will  
find that film is best for producing sensational large area color pictures, and the CCD is best for  
planets, small faint objects, and general scientific work such as variable star monitoring and  
position determination.  
2.2. How CCD Detectors Work  
The basic function of the CCD detector is to convert an incoming photon of light to an electron  
which is stored in the detector until it is read out, thus producing data which your computer  
can display as an image. It doesn't have to be displayed as an image. It could just as well be  
displayed as a spreadsheet with groups of numbers in each cell representing the number of  
electrons produced at each pixel. These numbers are displayed by your computer as shades of  
gray for each pixel site on your screen thus producing the image you see. How this is  
accomplished is eloquently described in a paper by James Janesick and Tom Elliott of the Jet  
Propulsion Laboratory:  
"Imagine an array of buckets covering a field. After a rainstorm, the buckets are  
sent by conveyor belts to a metering station where the amount of water in each  
bucket is measured. Then a computer would take these data and display a  
picture of how much rain fell on each part of the field. In a CCD the  
"raindrops" are photons, the "buckets" the pixels, the "conveyor belts" the CCD  
shift registers and the "metering system" an on-chip amplifier.  
Technically speaking the CCD must perform four tasks in generating an image.  
These functions are 1) charge generation, 2) charge collection, 3) charge transfer,  
and 4) charge detection. The first operation relies on a physical process known  
as the photoelectric effect - when photons or particles strikes certain materials  
free electrons are liberated...In the second step the photoelectrons are collected in  
the nearest discrete collecting sites or pixels. The collection sites are defined by  
an array of electrodes, called gates, formed on the CCD. The third operation,  
charge transfer, is accomplished by manipulating the voltage on the gates in a  
systematic way so the signal electrons move down the vertical registers from  
one pixel to the next in a conveyor-belt like fashion. At the end of each column  
is a horizontal register of pixels. This register collects a line at a time and then  
Page 5  
Section 2 - Introduction to CCD Cameras  
transports the charge packets in a serial manner to an on-chip amplifier. The  
final operating step, charge detection, is when individual charge packets are  
converted to an output voltage. The voltage for each pixel can be amplified off-  
chip and digitally encoded and stored in a computer to be reconstructed and  
displayed on a television monitor."1  
Readout Register  
Output  
Y=1  
Amplifier  
Y=N  
X=1  
X=M  
Figure 2.1 - CCD Structure  
2.2.1. Full Frame and Frame Transfer CCDs  
In the ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E, the CCD is read out electronically by shifting  
each row of pixels into a readout register at the Y=0 position of the CCD (shown in Figure 2.1),  
and then shifting the row out through an amplifier at the X=0 position. The entire array shifts  
up one row when a row is shifted into the readout register, and a blank row is inserted at the  
bottom. The electromechanical shutter built into the camera covers the CCD during the  
readout to prevent streaking of the image. Without a shutter the image would be streaked due  
to the fact that the pixels continue to collect light as they are being shifted out towards the  
readout register. CCDs with a single active area are called Full Frame CCDs.  
For reference, the ST-5C, ST-237, STV and ST-6 CCD cameras use a different type of  
CCD which is known as a Frame Transfer CCD. In these devices all active pixels are shifted  
very quickly into a pixel array screened from the light by a metal layer, and then read out. This  
makes it possible to take virtually streak-free images without a shutter. This feature is typically  
called an electronic shutter.  
2.3.  
Camera Hardware Architecture  
This section describes the ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E 2 CCD cameras from a  
systems standpoint. It describes the elements that comprise a CCD camera and the functions  
they provide. Please refer to Figure 2.2 below as you read through this section.  
1
"History and Advancements of Large Area Array Scientific CCD Imagers", James Janesick, Tom  
Elliott. Jet Propulsion Laboratory, California Institute of Technology, CCD Advanced Development  
Group.  
The ST-1001E does not have a second CCD for tracking.  
2
Page 6  
Section 2 - Introduction to CCD Cameras  
TE Cooler  
Clock  
Trackin g CCD  
Micro-  
con troller  
Drivers  
PC  
In terface  
16 Bit  
A/ D  
Pream p  
Telescope  
In terface  
Sh u tter  
Im agin g CCD  
Desktop  
Power  
Su pply  
Host Com pu t er  
Parallel In terface  
Figure 2.2 - CCD System Block Diagram  
As you can see from Figure 2.2, the ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E are completely  
self contained. Unlike our previous products, the ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E  
contain all the electronics in the optical head. There is no external CPU like the ST-5C, ST-237,  
ST-6 and STV.  
At the "front end" of any CCD camera is the CCD sensor itself. As we have already  
learned, CCDs are a solid state image sensor organized in a rectangular array of regularly  
spaced rows and columns. The ST-7E, ST-8E, ST-9E and ST-10E use two CCDs, one for  
imaging (Kodak KAF series) and one for tracking (TI TC211, like the ST-4/4X). Table 2.1 below  
lists some interesting aspects of the CCDs used in the various SBIG cameras.  
Array  
Number of  
Pixels  
Camera  
Tracking CCD  
ST-5C  
ST-237  
STV  
ST-6  
ST-7E  
ST-8E  
ST-9E  
CCD  
Dimensions  
2.6 x 2.6 mm  
3.2 x 2.4 mm  
4.7 x 3.6 mm  
4.7 x 3.0 mm  
8.6 x 6.5 mm  
Pixel Sizes  
13.75 x 16 µ  
10 x 10 µ  
7.4 x 7.4 µ  
14.8 x 14.8 µ  
23 x 27 µ  
TC211  
TC255  
TC237  
TC237  
TC241  
192 x 164  
320 x 240  
640 x 480  
320 x 200  
375 x 242  
765 x 510  
1530 x 1020 9 x 9 µ  
512 x 512  
2184 x 1472 6.8 x 6.8 µ  
1024 x 1024 24 x 24 µ  
KAF0401E 6.9 x 4.6 mm  
KAF1602E 13.8 x 9.2 mm  
KAF0261E 10.2 x 10.2 mm  
KAF3200E 14.9 x 10.0 mm  
KAF1001E 24.6 x 24.6 mm  
9 x 9 µ  
20 x 20 µ  
ST-10E  
ST-1001E  
Table 2.1 - Camera CCD Configurations  
Page 7  
Section 2 - Introduction to CCD Cameras  
The CCD is cooled with a solid-state a thermoelectric (TE) cooler. The TE cooler pumps heat  
out of the CCD and dissipates it into a heat sink which forms part of the optical head's  
mechanical housing. In the ST-7E and ST-8E cameras this waste heat is dumped into the air  
using passive radiators and a small fan, making the design and operation of the heads simple  
and not inconvenienced by requirements for liquid recirculation cooling. The ST-9E and ST-  
10E include SBIG's secondary TE/Liquid cooling booster.  
Since the CCD is cooled below 0°C, some provision must be made to prevent frost from  
forming on the CCD. The ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E have the CCD/TE Cooler  
mounted in a windowed hermetic chamber sealed with an O-Ring. The hermetic chamber does  
not need to be evacuated, another "ease of use" feature we employ in the design of our  
cameras. Using a rechargeable desiccant in the chamber keeps the humidity low, forcing the  
dew point below the cold stage temperature.  
Other elements in the self contained ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E include  
the preamplifier and an electromechanical shutter. The shutter makes taking dark frames a  
simple matter of pushing a button on the computer and provides streak-free readout. Timing  
of exposures in ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E cameras is controlled by this shutter.  
The Clock Drivers and Analog to Digital Converter interface to the CCD. The Clock  
Drivers convert the logic-level signals from the microcontroller to the voltage levels and  
sequences required by the CCD. Clocking the CCD transfers charge in the array and is used to  
clear the array or read it out. The Analog to Digital Converter (A/D) digitizes the data in the  
CCD for storage in the Host Computer.  
The microcontroller is used to regulate the CCD's temperature by varying the drive to  
the TE cooler. The external Power Supply provides +5V and ±12V to the cameras. Finally, the  
cameras contain a TTL level telescope interface port to control the telescope and the optional  
CFW-6A motorized color filter wheel.  
Although not part of the CCD Camera itself, the Host Computer and Software are an  
integral part of the system. SBIG provides software for the ST-7E, ST-8E, ST-9E, ST-10E and ST-  
1001E cameras for the IBM PC and Compatible computers (all cameras, DOS and Windows)  
and the ST-7E/8E are also supported by the Macintosh. The software allows image acquisition,  
image processing, and auto guiding with ease of use and professional quality. Many man-years  
and much customer feedback have gone into the SBIG software and it is unmatched in its  
capabilities.  
2.4. CCD Special Requirements  
This section describes the unique features of CCD cameras and the special requirements that  
CCD systems impose.  
2.4.1. Cooling  
Random readout noise and noise due to dark current combine to place a lower limit on  
the ability of the CCD to detect faint light sources. SBIG has optimized the ST-7E, ST-8E, ST-  
9E, ST-10E and ST-1001E to achieve readout noises below 20 electrons rms for two reads (light -  
dark). This will not limit most users. The noise due to the dark current is equal to the square  
root of the number of electrons accumulated during the integration time. For these cameras,  
the dark current is not significant until it accumulates to more than 280 electrons. Dark current  
is thermally generated in the device itself, and can be reduced by cooling. All CCDs have dark  
Page 8  
Section 2 - Introduction to CCD Cameras  
current, which can cause each pixel to fill with electrons in only a few seconds at room  
temperature even in the absence of light. By cooling the CCD, the dark current and  
corresponding noise is reduced, and longer exposures are possible. In fact, for roughly every 5  
to 6° C of additional cooling, the dark current in the CCD is reduced to half. The ST-7E and ST-  
8E have a single stage TE cooler and a temperature sensing thermistor on the CCD mount to  
monitor the temperature. The ST-1001E has two-stage cooling. The ST-9E and ST-10E have a  
supplemental second stage cooling booster with water cooling as an option (described in  
section 6.1). The same cooling booster used on the ST-9E and ST-10E may be added to the ST-  
7E, ST-8E or ST-1001E. The microcontroller controls the temperature at a user-determined  
value for long periods. As a result, exposures hours long are possible, and saturation of the  
CCD by the sky background typically limits the exposure time. At 0 °C the dark current in the  
ST-7E, ST-8E and ST-10E, high-resolution mode, is only 60 electrons per minute! The ST-1001E  
and ST-9E, with bigger pixels, have roughly 8 to 15 times this amount of dark current,  
respectively, due largely to the larger pixel area but also due to the inherent higher bulk dark  
current in the devices. That's why we include the cooling booster with the ST-9E and two-stage  
cooling plus the option of an additional cooling booster for the ST-1001E.  
The sky background conditions also increase the noise in images, and in fact, as far as  
the CCD is concerned, there is no difference between the noise caused by dark current and that  
from sky background. If your sky conditions are causing photoelectrons to be generated at the  
-
rate of 100 e /pixel/sec, for example, increasing the cooling beyond the point where the dark  
current is roughly half that amount will not improve the quality of the image. This very reason  
is why deep sky filters are so popular with astrophotography. They reduce the sky  
background level, increasing the contrast of dim objects. They will improve CCD images from  
very light polluted sights.  
2.4.2. Double Correlated Sampling Readout  
During readout, the charge stored in a pixel is stored temporarily on a capacitor. This capacitor  
converts the optically generated charge to a voltage level for the output amplifier to sense.  
When the readout process for the previous pixel is completed, the capacitor is drained and the  
next charge shifted, read, and so on. However, each time the capacitor is drained, some  
residual charge remains.  
This residual charge is actually the dominant noise source in CCD readout electronics.  
This residual charge may be measured before the next charge is shifted in, and the actual  
difference calculated. This is called double correlated sampling. It produces more accurate  
data at the expense of slightly longer read out times (two measurements are made instead of  
one). The ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E utilize double correlated sampling to  
-
-
produce the lowest possible readout noise. At 11e to 16e rms per read these cameras are  
unsurpassed in performance.  
2.4.3. Dark Frames  
No matter how much care is taken to reduce all sources of unwanted noise, some will remain.  
Fortunately, however, due to the nature of electronic imaging and the use of computers for  
storing and manipulating data, this remaining noise can be drastically reduced by the  
subtraction of a dark frame from the raw light image. A dark frame is simply an image taken  
Page 9  
Section 2 - Introduction to CCD Cameras  
at the same temperature and for the same duration as the light frame with the source of light to  
the CCD blocked so that you get a "picture" of the dark. This dark frame will contain an image  
of the noise caused by dark current (thermal noise) and other fixed pattern noise such as read  
out noise. When the dark frame is subtracted from the light frame, this pattern noise is  
removed from the resulting image. The improvement is dramatic for exposures of more than a  
minute, eliminating the many "hot" pixels one often sees across the image, which are simply  
pixels with higher dark current than average.  
2.4.4. Flat Field Images  
Another way to compensate for certain unwanted optical effects is to take a "flat field image"  
and use it to correct for variations in pixel response uniformity across the area of your dark-  
subtracted image. You take a flat field image of a spatially uniform source and use the  
measured variations in the flat field image to correct for the same unwanted variations in your  
images. The Flat Field command allows you to correct for the effects of vignetting and  
nonuniform pixel responsivity across the CCD array.  
The Flat Field command is very useful for removing the effects of vignetting that may  
occur when using a field compression lens and the fixed pattern responsivity variations present  
in all CCDs. It is often difficult to visually tell the difference between a corrected and  
uncorrected image if there is little vignetting, so you must decide whether to take the time to  
correct any or all of your dark-subtracted images. It is always recommended for images that  
are intended for accurate photometric measurements.  
Appendix D describes how to take a good flat field. It's not that easy, but we have  
found a technique that works well for us.  
2.4.5. Pixels vs. Film Grains  
Resolution of detail is determined, to a certain degree, by the size of the pixel in the detector  
used to gather the image, much like the grain size in film. The pixel size of the detector in the  
ST-10E is 6.8 x6.8 microns (1 micron = 0.001mm, 0.04 thousandths of an inch). In the ST-7E  
and ST-8E it is 9 x 9 microns, in the ST-9E it's 20 x 20 microns and in the ST-1001E it is 24 x 24  
microns. However, the effects of seeing are usually the limiting factor in any good photograph  
or electronic image. On a perfect night with excellent optics an observer might hope to achieve  
sub-arcsecond seeing in short exposures, where wind vibration and tracking error are minimal.  
With the average night sky and good optics, you will be doing well to achieve stellar images in  
a long exposure of 3 to 6 arcseconds halfwidth. This will still result in an attractive image,  
though.  
Using an ST-7E or ST-8E camera with their 9 micron pixels, an 8" f/10 telescope will  
produce a single pixel angular subtense of 0.9 arcsecond. An 8" f/4 telescope will produce  
images of 2.5 arcseconds per pixel. If seeing affects the image by limiting resolution to 6  
arcseconds, you would be hard pressed to see any resolution difference between the two focal  
lengths as you are mostly limited by the sky conditions. However, the f/4 image would have a  
larger field of view and more faint detail due to the faster optic. The ST-9, with its 20 micron  
pixels would have the same relationship at roughly twice the focal length or a 16 inch f/10  
telescope. See table 4.4 for further information.  
A related effect is that, at the same focal length, larger pixels collect more light from  
nebular regions than small ones, reducing the noise at the expense of resolution. While many  
Page 10  
Section 2 - Introduction to CCD Cameras  
people think that smaller pixels are a plus, you pay the price in sensitivity due to the fact that  
smaller pixels capture less light. For example, the ST-9E with its large 20 x 20 micron pixels  
captures five times as much light as the ST-7E and ST-8E's 9 micron square pixels. For this  
reason we provide 2x2 or 3x3 binning of pixels on most SBIG cameras. With the ST-7 and ST-8,  
for instance, the cameras may be configured for 18 or 27-micron square pixels. Binning is  
2
selected using the Camera Setup Command. It is referred to as resolution (High = 9µ pixels,  
2
2
Medium = 18µ pixels, Low = 27µ pixels). When binning is selected the electronic charge from  
groups of 2x2 or 3x3 pixels is electronically summed in the CCD before readout. This process  
adds no noise and may be particularly useful on the ST-10E with its very small 6.8 micron  
pixels. Binning should be used if you find that your stellar images have a halfwidth of more  
than 3 pixels. If you do not bin, you are wasting sensitivity without benefit. Binning also  
shortens the download time.  
The halfwidth of a stellar image can be determined using the crosshairs mode. Find the  
peak value of a relatively bright star image and then find the pixels on either side of the peak  
where the value drops to 50% of the peak value (taking the background into account, if the star  
is not too bright). The difference between these pixel values gives the stellar halfwidth.  
Sometimes you need to interpolate if the halfwidth is not a discrete number of pixels.  
Another important consideration is the field of view of the camera. For comparison, the  
diagonal measurement of a frame of 35mm film is approximately 43mm, whereas the diagonal  
dimension of the ST-7E chip is approximately 8 mm. The relative CCD sizes for all of the SBIG  
cameras and their corresponding field of view in an 8" f/10 telescope are given below:  
Camera  
Tracking CCD  
ST-5C  
ST-237  
STV  
ST-6  
ST-7E  
ST-8E  
ST-9E  
Array Dimensions  
2.64 x 2.64 mm  
3.20 x 2.40 mm  
4.74 x 3.55 mm  
4.74 x 2.96 mm  
8.63 x 6.53 mm  
6.89 x 4.59 mm  
13.8 x 9.18 mm  
10.2 x 10.2 mm  
14.9 x 10.0 mm  
24.6 x 24.6 mm  
36 x 24 mm  
Diagonal  
3.73 mm  
4.00 mm  
5.92 mm  
5.58 mm  
10.8 mm  
8.28 mm  
16.6 mm  
14.4 mm  
17.9 mm  
34.8 mm  
43 mm  
Field of View at 8" f/10  
4.5 x 4.5 arcminutes  
5.6 x 4.2 arcminutes  
8.2 x 6.1 arcminutes  
8.2 x 5.1 arcminutes  
14.6 x 11 arcminutes  
11.9 x 7.9 arcminutes  
23.8 x 15.8 arcminutes  
17.6 x 17.6 arcminutes  
25.1 x 16.9 arcminutes  
41.5 x 41.5 arcminutes  
62 x 42 arcminutes  
ST-10E  
ST-1001E  
35mm  
Table 2.2 - CCD Array Dimensions  
2.4.6. Guiding  
Any time you are taking exposures longer than several seconds, whether you are using a film  
camera or a CCD camera, the telescope needs to be guided to prevent streaking. While modern  
telescope drives are excellent with PEC or PPEC, they will not produce streak-free images  
without adjustment every 30 to 60 seconds. The ST-7E, ST-8E, ST-9E and ST-10E allow  
simultaneous guiding and imaging, called self-guiding (US Patent 5,525,793). This is possible  
because of the unique design employing 2 CCDs. One CCD guides the telescope while the  
other takes the image. This resolves the conflicting requirements of short exposures for  
guiding accuracy and long exposures for dim objects to be met, something that is impossible  
with single CCD cameras. Up to now the user either had to set up a separate guider or use  
Page 11  
Section 2 - Introduction to CCD Cameras  
Track and Accumulate to co-add several shorter images. The dual CCD design allows the  
guiding CCD access to the large aperture of the main telescope without the inconvenience of  
off-axis radial guiders. Not only are guide stars easily found, but the problems of differential  
deflection between guide scope and main scope eliminated. Due to the large size of the  
imaging CCD in the ST-1001E, however, a second CCD for tracking cannot be used. For this  
model camera we recommend the STV autoguider.  
Track and Accumulate is another SBIG patented process (US #5,365,269) whereby short  
exposures are taken and added together with appropriate image shifts to align the images. It is  
supported by the ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E camera software, but will generally  
not produce as good as results as self guiding, where the corrections are more frequent and the  
accumulated readout noise less. It is handy when no connection to the telescope drive is  
possible and also works best on cameras with larger pixels like the ST-1001E. For cameras with  
smaller pixels such as the ST-7, ST-8 and ST-10E, SBIG is proud to make self-guiding available  
to the amateur, making those long exposures required by the small pixel geometry easy to  
achieve!  
2.5. Electronic Imaging  
Electronic images resemble photographic images in many ways. Photographic images are  
made up of many small particles or grains of photo sensitive compounds which change color  
or become a darker shade of gray when exposed to light. Electronic images are made up of  
many small pixels which are displayed on your computer screen to form an image. Each pixel  
is displayed as a shade of gray, or in some cases a color, corresponding to a number which is  
produced by the electronics and photo sensitive nature of the CCD camera. However,  
electronic images differ from photographic images in several important aspects. In their most  
basic form, electronic images are simply groups of numbers arranged in a computer file in a  
particular format. This makes electronic images particularly well suited for handling and  
manipulation in the same fashion as any other computer file.  
An important aspect of electronic imaging is that the results are available immediately.  
Once the data from the camera is received by the computer, the resulting image may be  
displayed on the screen at once. While Polaroid cameras also produce immediate results,  
serious astrophotography ordinarily requires hypersensitized or cooled film, a good quality  
camera, and good darkroom work to produce satisfying results. The time lag between  
exposure of the film and production of the print is usually measured in days. With electronic  
imaging, the time between exposure of the chip and production of the image is usually  
measured in seconds.  
Another very important aspect of electronic imaging is that the resulting data are  
uniquely suited to manipulation by a computer to bring out specific details of interest to the  
observer. In addition to the software provided with the camera, there are a number of  
commercial programs available which will process and enhance electronic images. Images  
may be made to look sharper, smoother, darker, lighter, etc. Brightness, contrast, size, and  
many other aspects of the image may be adjusted in real time while viewing the results on the  
computer screen. Two images may be inverted and electronically "blinked" to compare for  
differences, such as a new supernova, or a collection of images can be made into a large mosaic.  
Advanced techniques such as maximum entropy processing will bring out otherwise hidden  
detail.  
Page 12  
Section 2 - Introduction to CCD Cameras  
Of course, once the image is stored on a computer disk, it may be transferred to another  
computer just like any other data file. You can copy it or send it via modem to a friend, upload  
it to your favorite bulletin board or online service, or store it away for processing and analysis  
at some later date.  
We have found that an easy way to obtain a hard copy of your electronic image is to  
photograph it directly from the computer screen. You may also send your image on a floppy  
disk to a photo lab which has digital photo processing equipment for a professional print of  
your file. Make sure the lab can handle the file format you will send them. Printing the image  
on a printer connected to your computer is also possible depending on your software/printer  
configuration. There are a number of software programs available, which will print from your  
screen. However, we have found that without specialized and expensive equipment, printing  
images on a dot matrix or laser printer yields less than satisfactory detail. However, if the  
purpose is simply to make a record or catalog the image file for easy identification, a dot matrix  
or laser printer should be fine. Inkjet printers are getting very good, though.  
2.6. Black and White vs. Color  
The first and most obvious appearance of a CCD image is that it is produced in shades of gray,  
rather than color. The CCD chip used in SBIG cameras itself does not discriminate color and  
the pixel values that the electronics read out to a digital file are only numbers proportional to  
the number of electrons produced when photons of any wavelength happen to strike its  
sensitive layers.  
Of course, there are color video cameras, and a number of novel techniques have been  
developed to make the CCD chip "see" color. The most common way implemented on  
commercial cameras is to partition the pixels into groups of three, one pixel in each triplet  
"seeing" only red, green or blue light. The results can be displayed in color. The overall image  
will suffer a reduction in resolution on account of the process. A newer and more complicated  
approach in video cameras has been to place three CCD chips in the camera and split the  
incoming light into three beams. The images from each of the three chips, in red, green and  
blue light is combined to form a color image. Resolution is maintained. For normal video  
modes, where there is usually plenty of light and individual exposures are measured in small  
fractions of a second, these techniques work quite well. However, for astronomical work,  
exposures are usually measured in seconds or minutes. Light is usually scarce. Sensitivity and  
resolution are at a premium. The most efficient way of imaging under these conditions is to  
utilize all of the pixels, collecting as many photons of any wavelength, as much of the time as  
possible.  
In order to produce color images in astronomy, the most common technique is to take  
three images of the same object using a special set of filters and then recombine the images  
electronically to produce a color composite or RGB color image. SBIG offers as an option a  
motorized color filter wheel. The accessory is inserted between the telescope and the CCD  
head. An object is then exposed using a red filter. The wheel is commanded to insert the green  
filter in place, and another image taken. Finally a blue image is taken. When all three images  
have been saved, they may be merged into a single color image using SBIG or third party color  
software.  
Page 13  
Section 3 - At the Telescope with a CCD Camera  
3.  
At the Telescope with a CCD Camera  
This section describes what goes on the first time you take your CCD camera out to the  
telescope. You should read this section throughout before working at the telescope. It will  
help familiarize you with the overall procedure that is followed without drowning you in the  
details. It is recommended you first try operating the camera in comfortable, well lit  
surroundings to learn its operation.  
3.1. Step by Step with a CCD Camera  
In the following sections we will go through the steps of setting up and using your CCD  
camera. The first step is attaching the camera to the telescope. The next step is powering up  
the camera and establishing a communication link to your computer. Then you will want to  
focus the system, find an object and take an image. Once you have your light image with a  
dark frame subtracted, you can display the image and process the results to your liking. Each  
of these steps is discussed in more detail below.  
3.2. Attaching the Camera to the Telescope  
ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E cameras are similar in configuration. The CCD head  
attaches to the telescope by slipping it into the eyepiece holder or attaching it via t-threads. A  
fifteen-foot cable runs from the head to the host computer's parallel port. The camera is  
powered by a desktop power supply. Operation from a car battery is possible using the  
optional 12V power supply or with a 12V to 110V power inverter.  
Connect the CCD head to the parallel port of your computer using the supplied cable  
and insert the CCD Camera's nosepiece into your telescope's eyepiece holder. Fully seat the  
camera against the end of the draw tube so that once focus has been achieved you can swap  
out and replace the camera without having to refocus. Orient the camera so that the CCD's  
axes are aligned in Right Ascension and Declination. Use Figure 3.1 below showing the back of  
the optical head as a guide for the preferred orientation. Any orientation will work, but it is  
aggravating trying to center objects when the telescope axes don't line up fairly well with the  
CCD axes.  
Next, connect the power cable and plug in the desktop power supply. The red LED on  
the rear of the camera should glow and the fan should spin indicating power has been applied  
to the unit. We recommend draping the cables over the finderscope, saddle or mount to  
minimize cable perturbations of the telescope, and guard against the camera falling out of the  
drawtube to the floor. We also recommend using the T-Ring attachments for connecting the  
camera to the telescope, as the cameras are heavy.  
Page 15  
Section 3 - At the Telescope with a CCD Camera  
RA  
DEC  
*
Figure 3.1 Orientation of the Optical Head Viewed from Back.  
(Pixel 1,1 is at the upper left in this view)  
3.3. Establishing a Communications Link  
When the CCDOPS program is initiated it will automatically attempt to establish a link to the  
camera. This involves identifying the type of CCD head. If the software is successful the  
"Link" field in the Status Window is updated to show the type of camera found. If the camera  
is not connected, powered up, or the LPT port setting has not yet been properly set, a message  
will be displayed indicating that the software failed to establish a link to the camera. If this  
happens, use the Communications Setup command in the Misc menu to configure the  
CCDOPS software for the LPT port you are using. Then use the Establish COM Link  
command in the Camera Menu to establish communications with the camera.  
Note: It is not necessary to have a camera connected to your computer to run the software and  
display images already saved onto disk. It is only necessary to have a camera connected  
when you take new images.  
Once the COM link has been established you may need to set the camera's setpoint  
temperature in the Camera Setup command. The ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E  
power up regulating to whatever temperature the CCD is at, which in this case will be the  
ambient temperature. Use the Camera Setup command and choose a setpoint temperature  
approximately 30°C below the ambient temperature. Type in the setpoint, set the temperature  
control to active, and hit ENTER.  
3.4. Focusing the CCD Camera  
Focusing a CCD camera can be a tedious operation, so a few hints should be followed. Before  
using the software to focus the camera the first time you should place a diffuser (such as scotch  
tape or ground glass) at the approximate location of the CCD's sensitive surface behind the  
eyepiece tube and focus the telescope on the moon, a bright planet or a distant street lamp.  
Page 16  
Section 3 - At the Telescope with a CCD Camera  
This preliminary step will save you much time in initially finding focus. The approximate  
distance behind the eyepiece tube for each of our CCD cameras is listed in Table 3.1 below:  
Camera  
ST-5C  
ST-237/A/STV  
ST-6  
ST-7E/8E/9E/ST-10E  
ST-1001E  
Distance  
~0.66 inch  
~0.68 inch  
~0.56 inch  
~0.92 inch  
~0.68 inch  
Diffuser  
Back Focus Distance  
from Table 3.1  
Table 3.1 - Camera Back Focus  
To achieve fine focus, insert the CCD head into the eyepiece tube, taking care to seat it,  
and then enter the CCDOPS FOCUS mode. The Focus command automatically displays  
successive images on the screen as well as the peak brightness value of the brightest object in  
the field of view. Point the telescope at a bright star. Center the star image in the CCD, and  
adjust the focus until the star image is a small as can be discerned. Next, move the telescope to  
a field of fainter stars that are dimmer so the CCD is not saturated. Further adjust the focus to  
maximize the displayed star brightness in counts and minimize the star diameter. This can be  
tedious. It helps considerably if a pointer or marker is affixed to the focus knob so you can  
rapidly return to the best focus once you've gone through it.  
An exposure of 1 to 3 seconds is recommended to smooth out some of the atmospheric  
effects. While you can use the Full frame mode to focus, the frame rate or screen update rate  
can be increased significantly by using Planet mode. In Planet mode the Focus command takes  
a full image and then lets you position a variable sized rectangle around the star. On  
subsequent images the Planet mode only digitizes, downloads, and displays the small area you  
selected. The increase in frame rate is roughly proportional to the decrease in frame size,  
assuming you are using a short exposure.  
The telescope focus is best achieved by maximizing the peak value of the star image.  
You should be careful to move to a dimmer star if the peak brightness causes saturation. The  
saturation levels of the various resolution modes are shown in Table 3.2 below. Another point  
you should also be aware of is that as you approach a good focus, the peak reading can vary by  
30% or so. This is due to the fact that as the star image gets small, where an appreciable  
percentage of the light is confined to a single pixel, shifting the image a half a pixel reduces the  
peak brightness as the star's image is split between the two pixels. The Kodak CCD pixels are  
so small that this is not likely to be a problem.  
Resolution  
High Res  
Saturation Counts  
~20,000 for ST-7E/8E ABG Cameras,  
~40,000 for ST-7E/8E Non ABG cameras,  
~50,000 for ST-10E Camera  
~65,000 for ST-9E/1001E Cameras  
Med/Low Res ~65,000 for ST-7E/8E/9E/10E/1001E  
Table 3.2 - Saturation Values  
Once the best focus is found, the focusing operation can be greatly shortened the second time  
by removing the CCD head, being careful not to touch the focus knob. Insert a high power  
Page 17  
Section 3 - At the Telescope with a CCD Camera  
eyepiece and slide it back and forth to find the best visual focus, and then scribe the outside of  
the eyepiece barrel. The next time the CCD is used the eyepiece should be first inserted into  
the tube to the scribe mark, and the telescope visually focused and centered on the object. At  
f/6 the depth of focus is only 0.005 inch, so focus is critical. An adapter may be necessary to  
allow the eyepiece to be held at the proper focus position. SBIG sells extenders for this  
purpose.  
3.5. Finding and Centering the Object  
Once best focus is achieved, we suggest using "Dim" mode to help center objects. This mode  
gives a full field of view, but reduces resolution in order to increase the sensitivity, and  
digitization and download rate. If you have difficulty finding an object after obtaining good  
focus, check to be sure that the head is seated at best focus, then remove the head and insert a  
medium or low power eyepiece. Being careful not to adjust the focus knob on the telescope,  
slide the eyepiece in or out until the image appears in good focus. Then visually find and  
center the object, if it is visible to the eye. If not, use your setting circles carefully. Then, re-  
insert the CCD head and use FOCUS mode with an exposure time of about ten seconds, if it is  
dim. Center the object using the telescope hand controls.  
Note: With a 10 second exposure, objects like M51 or the ring nebula are easily detected with  
modest amateur telescopes. The cores of most galactic NGC objects can also be seen.  
3.6. Taking an Image  
Take a CCD image of the object by selecting the Grab command and setting the exposure time.  
Start out with the Image size set to full and Auto Display and Auto contrast enabled. The  
camera will expose the CCD for the correct time, and digitize and download the image. One  
can also take a dark frame immediately before the light image using the Grab command.  
Because the ST-7E/ST-8E/ST-9E have regulated temperature control, you may prefer to  
take and save separate dark images, building up a library at different temperatures and  
exposure times, and reusing them on successive nights. At the start it's probably easiest to just  
take the dark frames when you are taking the image. Later, as you get a feel for the types of  
exposures and setpoint temperatures you use, you may wish to build this library of dark  
frames.  
3.7. Displaying the Image  
The image can be displayed on the computer screen using the graphics capability of your PC.  
Auto contrast can be selected and the software will pick background and range values which  
are usually good for a broad range of images or the background and range values can be  
optimized manually to bring out the features of interest.  
The image can also be displayed as a negative image, or can be displayed with  
smoothing to reduce the graininess. Once displayed, the image can be analyzed using  
crosshairs, or can be cropped or zoomed to suit your tastes.  
Page 18  
Section 3 - At the Telescope with a CCD Camera  
3.8. Processing the Image  
If not done already, images can be improved by subtracting off a dark frame of equal exposure.  
You will typically do this as part of the Grab command although it can also be done manually  
using the Dark Subtract command. By subtracting the dark frame, pixels which have higher  
dark current than the average, i.e., "hot" pixels, are greatly suppressed and the displayed  
image appears much smoother. Visibility of faint detail is greatly improved.  
The CCDOPS program also supports the use of flat field frames to correct for vignetting  
and pixel to pixel variations, as well as a host of other image processing commands in the  
Utility menu. You can smooth or sharpen the image, flip it to match the orientation of  
published images for comparison, or remove hot or cold pixels.  
3.9. Advanced Capabilities  
The following sections describe some of the advanced features of SBIG cameras. While you  
may not use these features the first night, they are available and a brief description of them is  
in order for your future reference.  
3.9.1. Crosshairs Mode (Photometry and Astrometry)  
Using the crosshair mode3 enables examination of images on a pixel by pixel basis for such  
measurements as Stellar and Diffuse Magnitude, and measurement of stellar positions. The 16  
bit accuracy of SBIG systems produces beautiful low-noise images and allows very accurate  
brightness measurements to be made. With appropriate filters stellar temperature can be  
measured.  
In the crosshair mode, you move a small cross shaped crosshair around in the image  
using the keyboard or the mouse. As you position the crosshair, the software displays the  
pixel value beneath the crosshair and the X and Y coordinates of the crosshair. Also shown is  
the average pixel value for a box of pixels centered on the crosshair. You can change the size of  
the averaging box from 3x3 to 31x31 pixels to collect all the energy from a star.  
3.9.2. Sub-Frame Readout in Focus  
The Focus command offers several frame modes for flexibility and increased frame throughput.  
As previously discussed, the Full frame mode shows the entire field of view of the CCD with  
the highest resolution, digitizing and displaying all pixels.  
The "Dim" mode offers the same field of view but offers higher frame rates by reducing  
the image's resolution prior to downloading. The resolution is reduced by combining a  
neighboring block of pixels into a "super pixel". This reduces the download and display times  
proportionately, as well as improving sensitivity. It is great for finding and centering objects.  
The Planet mode is suggested if high spatial resolution is desired for small objects like  
planets. The Planet mode allows you to select a small sub-area of the entire CCD for image  
acquisition. The highest resolution is maintained but you don't have to waste time digitizing  
and processing pixels that you don't need. Again, the image throughput increase is  
proportional to the reduction in frame size. It can be entered from Auto mode.  
3
Under DOS the Crosshairs mode is accessed through the Display Image command in the Analysis  
mode.  
Page 19  
Section 3 - At the Telescope with a CCD Camera  
Another aspect of the Focus command and its various modes is the Camera Resolution4  
setting in the Camera Setup command. Briefly, the Resolution setting allows trading off image  
resolution (pixel size) and image capture time while field of view is preserved. High resolution  
with smaller pixels takes longer to digitize and download than Low resolution with larger  
pixels. The cameras support High, Medium, Low and Auto resolution modes. The Auto mode  
is optimized for the Focus command. It automatically switches between Low resolution for  
Full frame mode to provide fast image acquisition, and High resolution for Planet mode to  
achieve critical focus. While Auto resolution is selected all images acquired using the GRAB  
command will be high resolution.  
3.9.3. Track and Accumulate  
An automatic Track and Accumulate mode (SBIG patented) is available in CCDOPS which  
simplifies image acquisition for the typical amateur with an accurate modern drive. These  
drives, employing PEC or PPEC technology and accurate gears, only need adjustment every 30  
to 120 seconds. With Track and Accumulate the software takes multiple exposures and  
automatically co-registers and co-adds them. The individual exposures are short enough such  
that drive errors are not objectionable and the accumulated image has enough integrated  
exposure to yield a good signal to noise ratio.  
Operationally the camera will take an exposure, determine the position of a preselected  
star, co-register and co-add the image to the previous image, and then start the cycle over  
again. Up to 64 images can be co-added, and the software even allows making telescope  
corrections between images to keep the object positioned in the field of view. The resulting  
exposure is almost as good as a single long exposure, depending on the exposure used and sky  
conditions. The great sensitivity of the CCD virtually guarantees that there will be a usable  
guide star within the imaging CCD's field of view. This feature provides dramatic  
performance for the amateur, enabling long exposures with minimal setup!  
3.9.4. Autoguiding and Self Guiding  
The CCDOPS software allows the ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E cameras to be  
used as autoguiders and self-guiders (except the ST-1001E) through the commands in the Track  
menu. While these systems are not stand-alone like the ST-4, but require a host computer, they  
can accurately guide long duration astrophotographs and CCD images with equal accuracy.  
Their sensitivity is much greater than an ST-4, and the video display easier to use.  
When functioning as an autoguider, the CCD camera repeatedly takes images of a  
guide star, measures the star's position to a fraction of a pixel accuracy, and corrects the  
telescope's position through the hand controller. While autoguiding alleviates the user of the  
tedious task of staring through an eyepiece for hours at a time, it is by no means a cure to  
telescope drive performance. All the things that were important for good manually guided  
exposures still exist, including a good polar alignment, rigid tubes that are free of flexure and a  
fairly good stable mount and drive corrector. Remember that the function of an auto guider is  
to correct for the small drive errors and long term drift, not to slew the telescope.  
4
The Resolution setting in the Camera Setup command combines pixels before they are digitized.  
This is referred to as on-chip binning and offers increases in frame digitization rates.  
Page 20  
Section 3 - At the Telescope with a CCD Camera  
One of the reasons that SBIG autoguiders are often better than human guiders is that,  
rather than just stabbing the hand controller to bump the guide star back to the reticule, it gives  
a precise correction that is the duration necessary to move the guide star right back to its  
intended position. It knows how much correction is necessary for a given guiding error  
through the Calibrate Track command. The Calibrate Track command, which is used prior to  
autoguiding, exercises the telescope's drive corrector in each of the four directions, measuring  
the displacement of a calibration star after each move. Knowing the displacement and the  
duration of each calibration move calibrates the drive's correction speed. Once that is known,  
the CCD tracker gives the drive corrector precise inputs to correct for any guiding error.  
When self guiding is selected by invoking the Self Guiding command under the Track  
Menu, the computer prompts the user for the exposure time for the tracking and imaging  
CCDs. Once these are entered, the computer takes and displays an image with the tracking  
CCD, and the user selects a guide star using the mouse. Guide stars that are bright, but not  
saturating, and isolated from other stars are preferred. Once the star is selected, the computer  
starts guiding the telescope. When the telescope corrections settle down (usually once the  
backlash is all taken up in the declination drive) the user starts the exposure by striking the  
space bar. The computer then integrates for the prescribed time while guiding the telescope,  
and downloads the image for display.  
A calibration star should be chosen that is relatively bright and isolated. The calibration  
software can get confused if another star of comparable brightness moves onto the tracking  
CCD during a move. The unit will self guide on much fainter stars. Tests at SBIG indicate that  
the probability of finding a usable guide star on the tracking CCD is about 95% at F/6.3, in  
regions of the sky away from the Milky Way. If a guide star is not found the telescope position  
should be adjusted, or the camera head rotated by a multiple of 90 degrees to find a guide star.  
We recommend that the user first try rotating the camera 180 degrees. Rotating the camera  
will require recalibration of the tracking function.  
3.9.5. Auto Grab  
The Auto Grab command allows you to take a series of images at a periodic interval and log  
the images to disk. This can be invaluable for monitoring purposes such as asteroid searches  
or stellar magnitude measurements. You can even take sub-frame images to save disk space if  
you don't need the full field of view.  
3.9.6. Color Imaging  
The field of CCD color imaging is relatively new but expanding rapidly. Since all SBIG  
cameras are equipped with monochromatic CCDs, discriminating only light intensity, not  
color, some provision must be made in order to acquire color images. SBIG offers a color filter  
wheel, the CFW-8, which provides this capability for the ST-7E, ST-8E, ST-9E and ST-10E.  
The color filter wheel allows remotely placing interference filters in front of the CCD in order to  
take multiple images in different color bands. These narrow band images are then combined to  
form a color image. With the SBIG system, a Red, Green and Blue filter are used to acquire  
three images of the object. The resulting images are combined to form a tri-color image using  
CCDOPS, CCDSoftV5 or third party software.  
Page 21  
Section 3 - At the Telescope with a CCD Camera  
Color imaging places some interesting requirements on the user that bear mentioning.  
First, many color filters have strong leaks in the infrared (IR) region of the spectrum, a region  
where CCDs have relatively good response. If the IR light is not filtered out then combining  
the three images into a color image can give erroneous results. If your Blue filter has a strong  
IR leak (quite common) then your color images will look Blue. For this reason, SBIG  
incorporates an IR blocking filter stack with the three color band filters.  
Second, since you have narrowed the CCD's wavelength response with the interference  
filters, longer exposures are required to achieve a similar signal to noise compared to what one  
would get in a monochrome image with wide spectral response. This is added to the fact that  
tri-color images require a higher signal to noise overall to produce pleasing images. With black  
and white images your eye is capable of pulling large area detail out of random noise quite  
well, whereas with color images your eye seems to get distracted by the color variations in the  
noisy areas of the image. The moral of the story is that while you can achieve stunning results  
with CCD color images, it is quite a bit more work.  
Page 22  
Section 4 - Camera Hardware  
4.  
Camera Hardware  
This section describes the modular components that make up the CCD Camera System and  
how they fit into the observatory, with all their connections to power and other equipment.  
4.1. System Components  
The ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E CCD cameras consist of four major components:  
the CCD Sensors and Preamplifier, the Readout/Clocking Electronics, the Microcontroller, and  
the power supply. All the electronics are packaged in the optical head in these cameras with  
an external desktop power supply.  
The CCDs, Preamplifier, and Readout Electronics are mounted in the front of the optical  
head. The optical head interfaces to the telescope through a 1.25 inch (or larger) draw tube,  
sliding into the telescope's focus mechanism. The placement of the preamplifier and readout  
electronics close to the CCD is necessary to achieve good noise performance. The  
Microcontroller is housed in the rear of the Optical Head along with the interface logic to the  
PC and Telescope.  
4.2. Connecting the Power  
The desktop power supply is designed to run off wall voltage, and both domestic (110 VAC)  
and European (220 VAC) versions are available. In the field however, battery operation may be  
the most logical choice. In that case you need to use the optional 12V power supply or a  
12VDC to 110 VAC power inverter.  
4.3. Connecting to the Computer  
The ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E CCD Cameras are supplied with a 15 foot cable  
to connect the system to the host computer. The connection is between the camera and the  
Host Computer's parallel LPT port. The camera will operate properly with up to 50 feet of  
cable to the computer. The computer's parallel port is easily identified since it is a 25 pin  
Female D connector.  
4.4. Connecting the Relay Port to the Telescope  
The ST-7E, ST-8E, ST-9E, ST-10E and ST-1001E camera systems can be used as autoguiders  
where the telescope's position is periodically corrected for minor variations in the RA and DEC  
drives. The host software functions as an autoguider in three modes: the Track mode, the SBIG  
patented Track and Accumulate mode, and the SBIG patented Self Guided mode (except for  
the ST-1001E).  
In the Track mode and Self Guided mode the host software corrects the telescope as  
often as once every 2 seconds to compensate for drift in the mount and drive system. The host  
software and the CCD camera operate in tandem to repeatedly take exposures of the  
designated guide star, calculate its position to a tenth of a pixel accuracy, and then  
automatically activate the telescope's controller to move the star right back to its intended  
position. It does this tirelessly to guide long duration astrophotographs.  
In the Track and Accumulate mode the software takes a series of images and  
automatically co-registers and co-adds the images to remove the effects of telescope drift.  
Page 23  
Section 4 - Camera Hardware  
Typically you would take ten 1 minute "snapshots" to produce an image that is comparable to a  
single 10 minute exposure except that no guiding is required. The reason no guiding is  
required is that with most modern telescope mounts the drift over the relatively short 1 minute  
interval is small enough to preserve round star images, a feat that even the best telescope  
mounts will not maintain over the longer ten minute interval. The Track and Accumulate  
software does allow correction of the telescope position in the interval between snapshots to  
keep the guide star grossly positioned within the field of view, but it is the precise co-  
registration of images that accounts for the streakless images.  
The host software and the CCD camera control the telescope through the 9-pin  
Telescope port on the camera. This port provides active low open collector signals to the  
outside world. By interfacing the camera to the telescope's controller the CPU is able to move  
the telescope as you would: by effectively closing one of the four switches that slews the  
telescope.  
Note: You only need to interface the camera's Telescope port to your telescope if you are  
planning on using the camera system as an autoguider or selfguider, or feel you need to  
have the Track and Accumulate command make telescope corrections between images  
because your drive has a large amount of long term drift.  
Some recent model telescopes (like the Celestron Ultima and the Meade LX200) have  
connectors on the drive controller that interface directly to the camera's TTL level Telescope  
port. All that's required is a simple cable to attach the 9 pin Telescope port to the telescope's  
telephone jack type CCD connector. SBIG includes its TIC-78 (Tracking Interface Cable  
Adapter) for this express purpose although it is easy to modify a standard 6-pin telephone  
cable for interface to the Telescope port (see Appendix A for specific pin outs, etc.). The TIC-78  
plugs into the 9-pin port on the ST-7E/8E/9E/10E/1001E, and a standard phone cable, which  
we supply, connects the adapter to the telescope drive. Note: phone cables come in a few  
variations. We use the six-pin cable, and the pin order is reversed left to right relative to the  
connector from one end to the other. This is identical to what is typically sold at Radio Shack  
stores as an extension cable.  
4.4.1 Using Mechanical Relays  
Older telescopes generally require modifying the hand controller to accept input from the  
camera's Telescope port. The difficulty of this task varies with the drive corrector model and  
may require adding external relays if your drive corrector will not accept TTL level signals. We  
maintain a database of instructions for the more popular telescopes that we will gladly share  
with you. For a minimal charge will also modify your hand controllers if you feel you do not  
have the skills necessary to accomplish such a task. We sell a mechanical relay box that  
interfaces to the ST-7E/8E/9E/10E/1001E, and will interface to the older drives. Contact SBIG  
for more information.  
In general, the camera has four signals that are used in tracking applications. There is  
one output line for each of the four correction directions on the hand controller (North, South,  
East and West). Our previous cameras had internal relays for the telescope interface, but with  
the proliferation of TTL input telescopes the relays were removed in the ST-7 and ST-8 (we do  
offer an external relay adapter accessory). The following paragraphs describe the general-  
purpose interface to the telescope which involves using external relays.  
Page 24  
Section 4 - Camera Hardware  
In our older camera models and in the optional relay adapter accessory, each of the  
relays has a Common, a Normally Open, and a Normally Closed contact. For example, when  
the relay is inactivated there is a connection between the Common and the Normally Closed  
contact. When the relay is activated (trying to correct the telescope) the contact is between the  
Common and the Normally Open contacts.  
If your hand controller is from a relatively recent model telescope it probably has four  
buttons that have a "push to make" configuration. By "push to make" we mean that the  
switches have two contacts that are shorted together when the button is pressed. If that's the  
case then it is a simple matter of soldering the Common and Normally Open leads of the  
appropriate relay to the corresponding switch, without having to cut any traces, as shown in  
Figure 4.1 below.  
B: Modified Push to Make Switch  
A: Unmodified Push to Make Switch  
common  
relay  
c
switch  
nc  
no  
switch  
normally open  
Figure 4.1 - Push to Make Switch Modification  
Another less common type of switch configuration (although it seems to have been used more  
often in older hand controllers) involve hand controller buttons that use both a push to make  
contact in conjunction with a push to break contact. The modification required for these  
switches involves cutting traces or wires in the hand controller. Essentially the relay's  
Normally Open is wired in parallel with the switch (activating the relay or pushing the hand  
controller button closes the Normally Open or Push to Make contact) while at the same time  
the Normally Closed contact is wired in series with the switch (activating the relay or pushing  
the hand controller button opens the Normally Closed or the Push to Break contact). This type  
of switch modification is shown in Figure 4.2 below.  
A: Unmodified Push to Make/ Break Switch  
B: Modified Push to Make/ Break Switch  
common  
c
common  
c
c
relay  
switch  
nc  
no  
nc  
no  
nc  
no  
normally open  
normally closed  
normally open  
normally closed  
Figure 4.2- Push to Make/Break Modification  
The last type of hand controller that is moderately common is the resistor joystick. In this  
joystick each axis of the joystick is connected to a potentiometer or variable resistor. Moving  
the joystick handle left or right rotates a potentiometer, varying the resistance between a  
central "wiper" contact and the two ends of a fixed resistor. The relays can be interfaced to the  
joystick as shown in Figure 4.3 below. Essentially the relays are used to connect the wire that  
Page 25  
Section 4 - Camera Hardware  
used to attach to the wiper to either end of the potentiometer when the opposing relays are  
activated.  
A
+ relay  
- relay  
c
c
A
wiper  
B
nc  
no  
nc  
no  
C
B
C
potentiometer  
B: Modified Joystick  
Figure 4.3 - Joystick Modification  
A: Unmodified Joystick  
A slight variation on the joystick modification is to build a complete joystick eliminator as  
shown in Figure 4.4 below. The only difference between this and the previous modification is  
that two fixed resistors per axis are used to simulate the potentiometer at its mid position. You  
do not need to make modifications to the joystick; you essentially build an unadjustable  
version. This may be easier than modifying your hand controller if you can trace out the  
wiring of your joystick to its connector.  
A
- relay  
+ relay  
c
c
A
nc  
no  
no  
nc  
wiper  
B
C
B
C
R
potentiometer  
R/ 2  
R/ 2  
B: Joystick Eliminator  
A: Unmodified Joystick  
Figure 4.4- Joystick Eliminator  
4.5. Modular Family of CCD Cameras  
With the introduction of the ST-6 CCD Camera in 1992 SBIG started a line of high quality, low  
noise, modular CCD cameras. The ST-7E, ST-8E and ST-9E were a second family of modular  
CCD cameras. The ST-10E and ST-1001E will also allow for upgrades to a faster USB interface  
when available.  
The benefits of a modular line of CCD Cameras are manyfold. Users can buy as much  
CCD Camera as they need or can afford, with the assurance that they can upgrade to higher  
performance systems in the future. With a modular approach, camera control software like  
CCDOPS can easily support all models. This last point assures a wide variety of third party  
Page 26  
Section 4 - Camera Hardware  
software. Software developers can produce one package for the many users across the model  
line instead of different packages for each of the cameras.  
While the SBIG cameras have many similarities, there are also important differences  
between the products. Table 4.2 below highlights the differences from a system's standpoint:  
A/D  
Temperature Electromechanical Electronic  
Camera  
Resolution Regulation  
Shutter/Shutter  
Wheel/Vane  
Shutter  
ST-5C  
ST-237A  
STV  
ST-6  
16 bits  
16 bits  
10+2 bits  
16 bits  
16 bits  
Closed Loop Shutter Wheel  
Closed Loop Shutter Wheel  
Closed Loop Shutter Wheel  
Closed Loop Vane  
0.01 second  
0.01 second  
0.001 second  
0.01 second  
None  
ST-7E/8E/9E/10E/1001E  
Closed Loop Shutter  
Table 4.2 - System Features  
How these features affect the average user are discussed in the paragraphs below:  
A/D Resolution - This is a rough indication of the camera's dynamic range. Higher precision  
A/D Converters are able to more finely resolve differences in light levels, or for  
larger CCDs with greater full well capacities, they are able to handle larger total  
charges with the same resolution.  
Temperature Regulation - In an open loop system like the original ST-4 the CCD cooling is  
either turned on or turned off. While this provides for adequate cooling of the  
CCD, the CCD's temperature is not regulated which makes it important to take  
dark frames in close proximity to the associated light frame. Closed loop  
systems regulate the CCD's temperature to an accuracy of ±0.1° C making dark  
frames useful over longer periods.  
Electromechanical Vane - Having the vane in the ST-6 means the host software can effectively  
"cover the telescope" and take dark frames remotely, without the user having to  
get up and physically cover the telescope.  
Electromechanical Shutter - Having the shutter in the ST-7E/8E/9E/10E/1001E gives streak-  
free readout and allows taking dark frames without having to cover the  
telescope. While the minimum exposure is 0.11 seconds, repeatability and area  
uniformity are excellent with SBIG's unique unidirectional shutter.  
Shutter Wheel - The Shutter Wheel, used in conjunction with the camera's Electronic shutter,  
allows you to cover the CCD for taking dark frames and in the case of the  
ST-5C/237/237A allows replacement with a mini internal color filter wheel.  
Electronic Shutter - Having an electronic shutter involves having a CCD with a frame transfer  
region. These CCDs actually have an array that has twice the number of rows  
advertised, where the bottom half is open to the light (referred to as the Image  
Area), and the top half is covered with a metalization layer (referred to as the  
Storage Area). In frame transfer CCDs at the end of the exposure, the pixel data  
from the Image Area is transferred into the Storage Area very rapidly where it  
can be read out with a minimum of streaking.  
In addition to the system level differences between the various cameras, Table 4.3 below  
quantifies the differences between different CCDs used in the cameras:  
Page 27  
Section 4 - Camera Hardware  
CCD Used Number of Pixel  
Array  
Dimension  
Read  
Noise  
-
Full Well  
Capacity  
-
Camera  
Pixels  
Dims.  
Tracking TC-211  
CCD  
TC-255  
ST-237A TC-237  
192 x 164  
13.75 x 16 µ 2.6 x 2.6mm  
12e rms 150Ke  
-
-
-
ST-5C  
320 x 240  
657 x 495  
320 x 200  
375 x 242  
10 x 10 µ  
7.4 x 7.4 µ  
14.8 x 14.8 µ 4.7 x 3.0mm  
23 x 27 µ  
9 x 9 µ  
3.2 x 2.4mm  
4.7 x 3.6mm  
20e rms 50Ke  
-
15e rms 20Ke  
-
STV  
ST-6  
ST-7E  
ST-8E  
TC-237  
TC-241  
KAF0401E 765 x 510  
20Ke-  
17e rms  
-
-
8.8 x 6.6mm  
6.9 x 4.6mm  
13.8 x 9.2mm 15e- rms  
30e rms 400Ke  
-
-5  
-4  
15e- rms  
50Ke /100Ke  
-
KAF1602E 1530 x 1020 9 x 9 µ  
50Ke /100Ke  
ST-9E  
ST-10E  
KAF0261E 512 x 512  
KAF3200E 2184 x 1472 6.8 x 6.8 µ  
20 x 20 µ  
10.2 x 10.2mm 13e- rms 180Ke-  
14.9 x 10.0mm 11e- rms 77Ke-  
24.6 x 24.6 mm 16e- rms 180Ke-  
ST-1001E KAF1001E 1024 x 1024 24 x 24 µ  
Table 4.3- CCD Differences  
How these various specifications affect the average user is described in the following  
paragraphs:  
Number of Pixels - The number of pixels in the CCD affects the resolution of the final images.  
The highest resolution device is best but it does not come without cost. Larger  
CCDs cost more money and drive the system costs up. They are harder to cool,  
require more memory to store images, take longer to readout, etc. With typical  
PC and Macintosh computer graphics resolutions, the CCDs used in the SBIG  
cameras offer a good trade off between cost and resolution, matching the  
computer's capabilities well.  
Pixel Dimensions - The size of the individual pixels themselves really plays into the user's  
selection of the system focal length. Smaller pixels and smaller CCDs require  
shorter focal length telescopes to give the same field of view that larger CCDs  
have with longer focal length telescopes. Smaller pixels can give images with  
higher spatial resolution up to a point. When the pixel dimensions (in  
arcseconds of field of view) get smaller than roughly half the seeing, decreasing  
the pixel size is essentially throwing away resolution. Another aspect of small  
pixels is that they have smaller full well capacities.  
For your reference, if you want to determine the field of view for a pixel  
or entire CCD sensor you can use the following formula:  
8.12x size (µm)  
Field of view (arcseconds) =  
focal length (inches)  
20.6x size(um)  
Field of view (arcseconds) =  
focal length(cm)  
where size is the pixel dimension or CCD dimension in millimeters and the focal  
5
The Kodak CCDs (KAF0400 and KAF1600) are available with or without Antiblooming Protection.  
Units with the Antiblooming Protection have one-half the full well capacity of the units without it.  
Page 28  
Section 4 - Camera Hardware  
length is the focal length of the telescope or lens. Also remember that 1° = 3600  
arcseconds.  
Read Noise - The readout noise of a CCD camera affects the graininess of short exposure  
images. For example, a CCD camera with a readout noise of 30 electrons will  
give images of objects producing 100 photoelectrons (very dim!) with a Signal to  
Noise (S/N) of approximately 3 whereas a perfect camera with no readout noise  
would give a Signal to Noise of 10. Again, this is only important for short  
exposures or extremely dim objects. As the exposure is increased you rapidly  
get into a region where the signal to noise of the final image is due solely to the  
exposure interval. In the previous example increasing the exposure to 1000  
photoelectrons results in a S/N of roughly 20 on the camera with 30 electrons  
readout noise and a S/N of 30 on the noiseless camera. It is also important to  
note that with the SBIG CCD cameras the noise due to the sky background will  
exceed the readout noise in 15 to 60 seconds on the typical amateur telescopes.  
Even the $30,000 priced CCD cameras with 10 electrons of readout noise will not  
produce a better image after a minute of exposure!  
Full Well Capacity - The full well capacity of the CCD is the number of electrons each pixel can  
hold before it starts to loose charge or bleed into adjacent pixels. Larger pixels  
hold more electrons. This gives an indication of the dynamic range the camera  
is capable of when compared to the readout noise, but for most astronomers this  
figure of merit is not all that important. You will rarely takes images that fill the  
pixels to the maximum level except for stars in the field of view. Low level  
nebulosity will almost always be well below saturation. While integrating  
longer would cause more build up of charge, the signal to noise of images like  
these is proportional to the square-root of the total number of electrons. To get  
twice the signal to noise you would have to increase the exposure 4 times. An  
-
ST-5C with its relatively low full well capacity of 50,000e could produce an  
image with a S/N in excess of 200!  
Antiblooming - Most SBIG CCD cameras have antiblooming protection. The TI CCDs used in  
the ST-5C, ST-237, ST-237A and ST-6 have it built into the CCDs and the Kodak  
CCDs used in the ST-7E and ST-8E have Antiblooming versions of the CCDs  
available. Blooming is a phenomenon that occurs when pixels fill up. As charge  
continues to be generated in a full pixel, it has to go somewhere. In CCDs  
without antiblooming protection the charge spills into neighboring pixels,  
causing bright streaks in the image. With the CCDs used in the SBIG cameras  
the excess charge can be drained off saturated pixels by applying clocking to the  
CCD during integration. This protection allows overexposures of 100-fold  
without blooming. The trade off is sensitivity. Antiblooming CCDs are less  
sensitive than non-antiblooming CCDs. In the case of the ST-7E and ST-8E, for  
example, the non-antiblooming versions are very roughly twice as sensitive.  
The CCDs used in the ST-9E, ST-10E and ST-1001E detectors do not come  
in an antiblooming version.  
From the telescope's point of view, the different models offer differing fields of view for a given  
focal length, or turned around, to achieve the same field of view the different models require  
differing focal lengths. Tables 4.4 and 4.5 below compare the fields of view for the cameras at  
several focal lengths, and vice versa.  
Page 29  
Section 4 - Camera Hardware  
LX200, 10" f/36  
C8, 8" f/10  
14" f/11  
Pixel  
Field of  
Pixel  
Size  
Field of  
View  
Pixel  
Size  
Field of  
View  
View  
Size  
Camera (arcmins)  
Tracking 4.2x4.2  
CCD  
(arcsecs) (arcmins)  
(arcsecs) (arcmins)  
(arcsecs)  
0.7x0.8  
1.3x1.5  
11.7x11.7  
3.7x4.3  
2.3x2.3  
ST-5C  
ST-237  
STV  
5.4x4.1  
8.0x6.0  
8.0x5.0  
14.6x11.1  
11.9x7.9  
23.8x15.8  
17.6x17.6  
25.1x16.9  
1.0x1.0  
14.4x10.8  
2.7x2.7  
2.0x2.0  
4.0x4.0  
6.2x7.3  
2.4x2.4  
2.4x2.4  
5.3x5.3  
1.8x1.8  
6.5x6.5  
2.8x2.1  
4.0x3.0  
4.0x2.5  
7.6x5.7  
6.1x4.1  
12.2x8.2  
9.1x9.1  
13.0x8.8  
21.6x21.6  
0.5x0.5  
0.75x0.75 21.6x16.2  
0.38x0.38  
0.75x0.75  
1.2x1.4  
0.5x0.5  
0.5x0.5  
1.1x1.1  
0.36x0.36  
1.27x1.27  
1.5x1.5  
2.3x2.7  
0.9x0.9  
0.9x0.9  
2.0x2.0  
0.7x0.7  
2.4x2.4  
21.6x13.5  
38.9x29.5  
31.2x20.8  
62.4x41.6  
46.2x46.2  
67.0x45.2  
111x111  
ST-6  
ST-7E  
ST-8E  
ST-9E  
ST-10E  
ST-1001E 41.6x41.6  
Table 4.4 - Field of View  
Focal Length to fill Focal Length to fill  
Object  
Moon  
Size  
0.5°  
Tracking CCD  
ST-7E  
275mm =  
11 inches  
760mm =  
30 inches  
Jupiter  
40 arcseconds  
13000mm =  
510 inches  
34000mm =  
1350 inches  
M51-Whirlpool  
Galaxy  
8 x 5  
arcminutes  
1040mm =  
41 inches  
3700mm =  
145 inches  
M27-Dumbell  
Nebula  
8.5 x 5.5  
arcminutes  
"
"
M57-Ring Nebula  
1.3 x 1  
arcminutes  
6400mm =  
250 inches  
23000mm =  
900 inches  
Table 4.5 - Focal Length Required  
From these numbers you can deduce that the popular C8, an 8" f/10 telescope will nicely frame  
many popular objects with the ST-7E whereas a much shorter system (f/3, perhaps achieved  
with a focal reducer) will frame the same objects for the tracking CCD. Another point to bear  
in mind is that, except for planetary images, you'll rarely take images where the pixel size in  
seconds of arc is down near the seeing limit. Most objects are relatively large, where the field  
of view is more important than whether the individual pixels are less than half the seeing.  
4.6 Connecting the older model CFW-6 filter wheel to the Camera  
The SBIG CFW-6 filter wheel can be connected to the camera using the relay port  
adapter provided with the camera. The adapter plugs into the 9 Pin connector on the  
6
f/6.3 with the Meade star digonal focal reducer.  
Page 30  
Section 4 - Camera Hardware  
ST-7E/8E/9E/10E head. The phone cable connection connects to the telescope drive to adjust  
the telescope position when collecting images using Self-Guiding, and the short cable with  
three wires connects to the CFW-6. The black wire of the three-wire group mates to the brown  
or black wire of the CFW-6.  
4.7 Battery Operation  
The ST-7E/8E/9E/10E can be operated off of a 12 volt car or marine battery using a the  
optional 12V power supply or using a power inverter. We have used the Radio Shack model  
22-132A, 12 volt DC to 115 volt AC Portable Power Inverter (140 watt) with good success. The  
camera draws 2.2 amps from the battery with this inverter, which should enable an evening's  
operation from a single battery. The cooling booster, if used, draws another 2 amps. We  
recommend a separate battery for the camera; using your vehicle's battery with the engine  
running may add undesirable readout noise, and using your vehicle's battery without the  
engine running may result in a long walk home in the dark!  
4.8 ST-1001E Differences  
The ST-7E/8E/9E/10E/1001E all use the same mechanical head. However, in the Model ST-  
1001E some modifications are made to accommodate the much larger imaging CCD.  
No Tracking CCD: All of the models mentioned above have a second CCD mounted in  
the head for tracking except the ST-1001E. For ST-1001E users who need to autoguide during  
exposures we recommend the STV stand alone autoguider. More information can be obtained  
at the phone number or e-mail address on the cover of this manual.  
AO-7 Compatibility: Because the AO-7 Adaptive Optics accessory uses the tracking CCD  
to determine corrections, the AO-7 cannot be used with the ST-1001E.  
CFW8 Filter Wheel Compatibility: Due to the large imaging CCD and corresponding  
larger optical path required for the ST-1001E, 1.25" filters such as those used in the CFW8 color  
filter wheel are too small for this camera. ST-1001E users may contact SBIG for  
recommendations of third party filters and accessories for photometric work or color imaging.  
Optical Center: The optical center of the window in the ST-1001E camera is moved  
slightly towards the center of the camera. ST-1001E users should not rely on the mechanical  
head drawings for the ST-7/8/9/10E cameras for this measurement. Please consult the SBIG  
web site or SBIG technical support for this measurement if needed.  
Two-stage Cooling with Optional Three-stage Cooling: The KAF-1001E CCD is nearly 1"  
square and requires more cooling capability than smaller CCDs to maintain a similar  
temperature. For this model camera an internal two-stage cooling system is standard. This  
requires a larger capacity power supply than the ST-7E/8E/9E/10E. In addition, the ST-1001E  
may be ordered with or retrofitted with the same optional Cooling Booster with water assist  
that is used on the other cameras giving three-stage cooling, if conditions require.  
12VDC Operation: The ST-1001E comes only with a 110VAC switching power supply  
that provides the correct voltages and current capacity required by the camera. An inverter  
should be used to operate from a 12VDC source such as a marine deep cycle battery.  
Page 31  
Section 5 - Advanced Imaging Techniques  
5.  
Advanced Imaging Techniques  
With practice, you will certainly develop methods of your own to get the most from your CCD  
camera. In this section we offer some suggestions to save you time getting started in each of  
the different areas outlined below, but these suggestions are by no means exhaustive.  
5.1. Lunar and Planetary Imaging  
When imaging the moon using most focal lengths available in astronomical telescopes, you will  
note that the moon's image typically fills the CCD. The image is also very bright. The best  
way to image the moon is to use neutral density filters to attenuate the light.  
You may also try an aperture mask to reduce the incoming light. If you feel that an  
aperture mask reduces resolution to an unacceptable degree, consider using a full aperture  
solar filter. This will result in an optimum exposure of only a few seconds. Another way to  
reduce the incoming light is to increase the effective focal length of your telescope by using a  
barlow lens or eyepiece projection. This is very desirable for planetary imaging since it also  
increases the image size.  
5.2. Deep Sky Imaging  
Ordinarily, with telescopes of 8" aperture or larger, a ten second exposure in focus mode, with  
a dark frame subtracted, will show most common deep sky objects except for the very faintest.  
This is a good starting point for finding and centering deep sky objects. If you find ten seconds  
to be insufficient, a one minute exposure will clearly show nearly any object you wish to image,  
particularly if 2x2 or 3x3 binning is selected. Using the Grab command, exposures of five  
minutes will generally give you a clear image with good detail. Of course, longer exposures  
are possible and desirable, depending on your telescope's tracking ability or your desire to  
guide. Once you have determined the longest exposure possible with your particular drive  
error, try Track and Accumulate exposures of a duration less than your measurable error. We  
have found that exposures of thirty seconds to two minutes are best depending on the focal  
length of telescope one is using. With the self guiding feature of the ST-7E/8E/9E/10E one  
can take long integrations while the internal tracking CCD guides the telescope.  
We highly recommend that you initially pursue deep sky imaging with a fast telescope,  
or focal reducer to produce a final F number of F/6.3 or faster at the CCD. The sensitivity  
advantage is considerable!  
5.3. Terrestrial Imaging  
An optional accessory for the SBIG cameras is the camera lens adapter. These accessories are  
made to accommodate most popular camera models. You may attach a camera lens in place of  
your telescope and use the CCD camera for very wide angle images of the night sky or for  
terrestrial views in daylight. Begin with a tenth second exposure at f/16 for scenes at normal  
room light and adjust as necessary for your conditions.  
5.4. Taking a Good Flat Field  
If you find that flat field corrections are necessary due to vignetting effects, CCD sensitivity  
variations, or for more accurate measurements of star magnitudes, try either taking an image of  
Page 33  
Section 5 - Advanced Imaging Techniques  
the twilight sky near the horizon or take an image of a blank wall or neutral grey card. The  
Kodak CCDs may have a low contrast grid pattern visible in the sky background. A flat field  
will eliminate this.  
Finding areas of the sky devoid of stars is very difficult after twilight. Therefore, you  
should take flat field images of the night sky after sunset, but long before you can see any stars.  
If this is not possible, take an image of a featureless wall or card held in front of the telescope.  
However, if using this second method, be sure that the wall or card is evenly illuminated.  
Appendix D describes how to do this. You will know if the flat field is good if the sky  
background in your images has little variation across the frame after flat fielding, displayed  
using high contrast (a range of 256 counts is good for showing this).  
If you plan on flat fielding Track and Accumulate images you should also refer to  
section 6.8. Since the same flat field is added to itself a number of times, be sure that you do  
not saturate the flat field image by starting with pixel values too high. Typically try to keep the  
pixel values between 10% to 20% of saturation for this purpose. For single flat field images, try  
to keep the values to approximately 50% of saturation.  
5.5. Building a Library of Dark Frames  
The ST-7E/8E/9E/10E/1001E have regulated temperature control, and therefore it is possible  
to duplicate temperature and exposure conditions on successive nights. You can set the  
camera TE cooler temperature to a value comfortably within reach on your average night, and  
then take and save on disk a library of dark frames for later use. This is a good project for a  
rainy night. We recommend you build a file of 5, 10, 20 ,40, and 60 minute dark frames at zero  
degrees Centigrade for a start. Otherwise you will find yourself wasting a clear night taking  
hour long dark frames!  
Note: Dark frames taken the same night always seem to work better. The adaptive  
dark subtract will help if the ambient temperature changes slightly.  
5.6. Changing the Camera Resolution  
The Camera Setup command allows you to select the resolution mode you wish to use for  
taking and displaying images. The ST-7E/8E/9E/10E/1001E cameras have High, Medium,  
Low and Auto modes. The High Resolution mode is the best for displaying the greatest detail  
since it utilizes the maximum number of pixels for your particular camera. The Medium  
Resolution Mode operates by combining 2x2 pixels giving the same field of view as High  
Resolution Mode, but with 1/4 the resolution. This results in significantly faster digitization  
and download times. Also, in Medium Resolution Mode, with larger pixels and comparable  
readout noise there is a better signal to noise ratio for very dim diffuse objects. This improved  
signal to noise ratio combined with faster digitization and download times makes Medium  
Resolution Mode ideal for finding and centering dim objects, and for imaging most objects.  
Additionally, a Low resolution mode is provided which bins the CCD 3x3 before readout.  
Low resolution mode is sensational for displaying faint nebulosity with short exposure times.  
In Auto Resolution Mode, the camera and software will always use High Resolution for all  
imaging and display functions except when you are in Full Frame Focus Mode. It will then  
automatically switch to Low Resolution Mode. If you further select Planet Mode for focusing,  
the camera will switch back to High Resolution on the selected box area. The small pixel size,  
Page 34  
Section 5 - Advanced Imaging Techniques  
is best for critical focusing. Planet mode will result in fast digitization and download times  
since only a small portion of the frame is read out.  
In general, you should pick a binning mode that yields stars with two to three pixels  
full width at half maximum. This is easily measured by using the crosshairs to determine the  
peak brightness of a relatively bright star, and determining the number of pixels between the  
50% values on either side of the peak. More than 3 pixels per stellar halfwidth merely wastes  
sensitivity without improved resolution.  
5.7. Flat Fielding Track and Accumulate Images  
This section gives the step by step procedure for flat field correcting images taken using the  
Track and Accumulate command.  
Flat field correcting images allow the user to remove the effects of CCD response non-  
uniformity (typically less than a few percent) and optical vignetting which for some optical  
systems can be as much as a 50% effect from center to edge. The CCDOPS software allows flat  
field correcting images using the Flat Field command, but some preparation must be made to  
use that command with Track and Accumulate images. Essentially you must prepare a special  
flat field correction image for Track and Accumulate images. This special preparation is  
necessary to have the same set of alignment and co-addition operations apply to the flat field  
file that have occurred in acquiring the Track and Accumulate image. In general, the following  
procedure should be followed when flat field correction of Track and Accumulate images is  
desired:  
1. Take a normal flat field image using the Grab command. You can use the dusk  
sky or a neutral gray or white card held in front of the telescope. Try to adjust  
the illumination and/or exposure so that the build up of light in the image  
yields values that when co-added several times will not overflow 65,000 counts.  
The number of times the image will need to be co-added without overflowing is  
set by how many snapshots you intend to use in Track and Accumulate. A  
good goal is to try and attain a maximum level in the flat field image of 1,000 to  
2,000 counts which will allow co-addition 32 times without overflow.  
Note: You will have to take a new flat field image anytime you change the optical  
configuration of your telescope such as removing and replacing the optical head in the  
eyepiece holder.  
2. Save the flat field image on your disk using the Save command. In the following  
discussions this flat field file will be referred to as FLAT.  
3. Take your Track and Accumulate image using the Track and Accumulate  
command and save it on the disk using the Save command. In the following  
discussions this Track and Accumulate image file will be referred to as IMAGE.  
4. Immediately after saving the IMAGE use the Save Track List command on the  
PC or activate the Track List window on the Mac and use the Save command to  
save the Track and Accumulate track list. The track list is a file that describes  
what alignment operations were done to the individual components of IMAGE  
to achieve the end result. In the following discussions this track list file will be  
referred to as TRACK.  
Page 35  
Section 5 - Advanced Imaging Techniques  
5. Repeat steps 3 and 4 as many times as desired for all the objects you wish to  
image, each time choosing a set of corresponding new names for the IMAGE  
and TRACK files.  
6. You will now create a combined flat field image for each Track and Accumulate  
image you captured. Invoke the Add by Track List command. The software  
will bring up a file directory dialog showing all the track list files. Select the  
TRACK file corresponding to the image you wish to correct. The software will  
load the TRACK file and present you with another file directory dialog showing  
all the images. Select the appropriate FLAT image. The software will align and  
co-add the FLAT image using the same operations it performed on the Track  
and Accumulate image. Finally save the combined flat field image using the  
Save command. In the following discussions this combined flat field image will  
be referred to as COMBINED-FLAT. Repeat this step for each of the TRACK  
files using a corresponding name for the COMBINED-FLAT image.  
7. You will now flat field correct the Track and Accumulate image with the  
combined flat field image. Use the Open command to load the IMAGE file, then  
use the Flat Field command. The software will present you with a file directory  
dialog where you should select the corresponding COMBINED-FLAT image.  
After the software has finished correcting the image you can view the results  
and save the flat field corrected image with the Save command. This image will  
be referred to as the CORRECTED-IMAGE file. Repeat this step for each of the  
IMAGE files using the corresponding COMBINED-FLAT image.  
At SBIG we have adopted the following naming convention for our various image and related  
files. If it helps you organize your files please feel free to adopt it or any method you feel helps  
sort out the process of naming files:  
Image type  
Name  
Uncorrected image (IMAGE)  
Flat field file (FLAT)  
Track list file (TRACK)  
Combined flat field (COMBINED-FLAT)  
Flat field corrected image (CORRECTED-IMAGE)  
XXXXXXXX.  
FLATXXXX.  
XXXXXXXX.TRK  
FLATXXXX.C  
XXXXXXXX.F  
(blank extension)  
(blank extension)  
5.8. Tracking Functions  
The CCDOPS software allows your ST-7E/8E/9E/10E/1001E to be used as an autoguider or  
self-guided (except ST-1001E) imager. It does not function as a stand-alone autoguider like the  
ST-4, but instead requires using a PC to perform the function. These cameras have  
considerably better sensitivity than the ST-4.  
CCD autoguiders alleviate you from having to stare down the eyepiece for hours at a  
time while guiding astrophotographs. They are not the end-all, cure-all approach to telescope  
mechanical problems, though. You still need a good polar alignment and a rigid mount  
between the guide scope and the main scope or you need to use an off-axis guider, with all its  
inherent difficulties. A good declination drive, free of backlash, is desirable although not  
Page 36  
Section 5 - Advanced Imaging Techniques  
absolutely necessary. Finally, modern drive correctors with periodic error correction (PEC) or  
permanent periodic error correction (PPEC) will ease the difficulty of achieving good results.  
The moral of the story is don't count on the CCD autoguider to fix all your problems.  
The better the drive, the better results you will obtain.  
Using the CCD as an autoguider requires interfacing the CPU's relay port to your hand  
controller (as discussed in section 4.4) and then training the CCD system on your telescope.  
This is done with the Calibrate Track command. Focus your system and then find and center  
on a moderately bright calibration star (1000 to 20,000 peak counts will do) without any nearby  
neighboring stars of similar brightness. Then execute Calibration command. It takes a  
sequence of five images. In the first image it determines the pixel position of the calibration  
star. In the four subsequent images, the software, in sequence, activates each of the telescope's  
four correction directions, measuring the displacement of the calibration star. From this  
calibration information, the software is able to calculate a precise correction when the guide star  
moves away from its intended position.  
At the start, you should pick a calibration star with roughly the same declination as the  
intended object since the telescope's correction speeds vary with declination. As you get used  
to the tracking functions you can calibrate on a star near the celestial equator and have the  
software adjust for different declinations for you.  
The tracking functions in CCDOPS are accessed through the Track menu. The Calibrate  
Track command, as described above, is used to calibrate your telescope's drive corrector. Once  
that is done, the Track command, which you would use for autoguiding astrophotographs,  
allows you to select a guide star in the field of view, and then repeatedly takes images,  
measuring the guide star's position and hence guiding error, and corrects the telescope.  
The Track and Accumulate command, discussed in section 3.9.2, also has provisions for  
making telescope corrections between images. This is necessary only if your drive has a large  
amount of long term drift, which results in Track and Accumulate images that are reduced in  
width.  
Finally, the Tracking Parameters command allows you to fine tune the CCDOPS  
tracking performance. You can deactivate the RA or DEC corrections or even deactivate both, a  
feature that can be used to monitor the uncorrected tracking accuracy of your drive. You can  
also fine tune the correction speeds if you find the telescope is consistently over or under  
correcting.  
Page 37  
Section 6 - Accessories for your CCD Camera  
6.  
Accessories for your CCD Camera  
This section briefly describes the different accessories available for your CCD camera.  
6.1. Cooling Booster  
The cooling booster, which is included with the ST-9E and ST-10E, is an option for the  
ST-7E, ST-8E and ST-1001E. It is a small module that gets installed inside the back  
compartment of the camera. This section tells you how to best use the Cooling Booster.  
Caution: Do not connect the power to the cooling booster unless the main  
power to the camera has already been connected and the fan is running. The  
fan is needed to keep the booster's heat sink cool.  
The cooling booster can be used either without the flowing water or with it. We do not  
recommend using water cooling in the field due to the extra equipment necessary such as  
water reservoir, pump, hoses and additional power required for the pump. For users who  
have fixed sites, or small observatories, the extra equipment required is easier to manage. The  
major problem one must deal with is routing the water tubes off the mount to minimize  
perturbations to the mount during tracking. In general, try to route tubes (and wires) over the  
mount, rather than just let the tubes dangle from the end of a long tube. Water-cooling is  
probably not necessary for most users when the air temperature is below 10 degrees C (50  
degrees F), since the dark current is fairly low already. Think of it as a summertime accessory!  
We do not recommend use of water cooling below freezing temperatures, where antifreeze  
must be added to the water. It is simply not necessary then. There is no problem with using  
the cooling booster with only air cooling in the winter, though.  
Without flowing water the cooling improvement is about 6 degrees C. With it the  
cooling improvement is about 15 degrees C. If you plan to use it without the water then you  
should disconnect the hoses from the camera and shake out the water trapped in the heat sink.  
Disconnecting the hoses will reduce the potential perturbation to your telescope mount.  
To operate the cooling booster without water cooling, mount the camera to the  
telescope as before and simply plug the auxiliary 12 volt supply jack into the connection on the  
camera back plate. Turn on the TE cooling to 100% by giving it a target temperature of –50  
degrees. After 10 minutes examine the camera temperature, and reset the set point to 3  
degrees C above the current temperature. This 3 degree temperature margin will enable the  
ST-7E/8E/9E to regulate the temperature accurately.  
To operate the camera with water cooling, the procedure is the same except that the  
water flow must be established before mounting the camera to the telescope, since the water  
pumps have limited pressure capability. To do this, put the camera at the same level as the  
water reservoir. Connect all the hoses, and make sure the water return goes back into the  
reservoir. Push the ¼ inch internal diameter (ID) hoses onto the nipples on the back of the  
camera so they seal. Attach one hose to the nipple onto the reducing connector that adapts the  
¼ inch ID hose to the ½ inch diameter hose from the pump.  
Turn on the pump, and let the flow establish itself through the hoses. Next, mount the  
camera to the telescope. If you always keep the return hose outlet near the reservoir level the  
pump will have no problem raising the water 2 meters (6 feet) off the floor. The limited  
pressure capacity of the pump is only a problem when you let the water fall back into the  
Page 39  
Section 6 - Accessories for your CCD Camera  
reservoir from a significant height above it, such a 0.3 meter (12 inches). Lastly, check for  
leaks!  
When using water cooling, avoid the temptation to put ice in the water to get the  
camera even colder. As the cooling booster is designed, the camera will not be cooled below  
ambient temperature if ambient temperature water is used. If colder water is used, the head  
may fog or frost up, depending on the dew point. . The exposed electronics inside the ST-7/8  
will get wet, and corrode. The hoses will start dripping condensation, and you will have a  
mess. Keep the ice for a cold drink!  
At the end of the evening, stop the pump, and raise the outlet hose above the camera to  
let all the water drain out of the system. Blowing it out with gently pressure helps clear the  
water. You can leave the hoses full of water, but if a leak occurs while you’re not there you  
may have a problem.  
When packing the camera for a long time, or at the end of summer, disconnect the  
hoses and blow out the heat sink to allow the enclosed spaces to dry out and minimize long  
term corrosion.  
6.2. Tri-color Imaging  
You can make splendid color images with your CCD camera by using the optional CFW-8  
color filter wheel. The CFW-8 attaches to the front of the CCD head and allows you to take  
images in red, green and blue light of the same object. When these images are aligned and  
processed a full color image results. (Note: Due to the large size of the imaging CCD in the ST-  
1001E, the CFW8 will not work with this model camera. A filter wheel with 2" filters is recommended).  
6.3. Camera Lens Adapters and Eyepiece Projection  
Camera Lens adapters are available for the ST-7E/8E/9E/10E/1001E. The camera lens adapter  
allows you to mount your camera lens in place of the telescope for very wide field views of the  
night sky or for daytime terrestrial imaging. The adapters are available for a variety of the  
most common camera lenses in use.  
The cameras themselves can be used for eyepiece projection assuming you already have  
a T-thread eyepiece projection adapter for. Instead of attaching a 35mm camera to the T-thread  
camera adapter you can attach the CCD camera by unscrewing the T-Thread nosepiece fom the  
front of the camera.  
6.4. Focal Reducers  
Several third party vendors, including Optec, Celestron and Meade, make focal reducers for  
their telescopes that could be used with the ST-7E/8E/9E/10E cameras. While most will work  
with the ST-7 some may have slight amounts of vignetting when used on the larger ST-  
8E/9E/10E. As of this writing we have not tested any focal reducers with the ST-1001E  
Note: Many very fast (F/3.3) commercial units will not accommodate a filter wheel.  
Page 40  
Section 6 - Accessories for your CCD Camera  
6.5. AO-7 and Lucy-Richardson Software  
The AO-7 is the world's only Adaptive Optics accessory for the amateur CCD market and it  
works only with the self guided feature of the ST-7E/8E/9E/10E. The AO-7 is essentially an  
electromechanical driven diagonal mirror that goes between the camera and the telescope.  
Using the AO-7 allows corrections to be made at rates up to 40 Hz. This yields perfect,  
backlash-free guiding and allows removal of some atmospheric effects. When used in  
conjunction with the CCDSharp software stellar peak brightness can double and stellar half-  
widths can shrink by 25%.  
6.6. SGS - Self-Guided Spectrograph  
The SGS Self Guided Spectrograph takes the tedium out of spectroscopy by allowing you to  
image and guide the source on the tracking CCD while acquiring its spectra on the imaging  
CCD. No more hunting around to place the object on the slit! With the SGS you can measure  
galactic redshifts, stellar classifications and determine nebular constituency. This is another  
example of the value of a self-guided camera like the ST-7E/8E/9E/10E.  
6.7. Third Party Products and Services  
There are numerous third party products and services available directed to the CCD user. We  
mention a few to give you an idea of the ever increasing interest in this new technology.  
Scanning through the pages of Sky & Telescope, CCD Astronomy, or Astronomy magazine  
reveals many third-party vendors of CCD accessories and software.  
6.7.1. Windows Software  
If you are a user interested in using Windows software with your camera, SBIG has CCDOPS  
for Windows. In addition, after January 2001, all SBIG cameras will include CCDSoftV5 which  
is a joint software development of SBIG and Software Bisque. There are also several other  
commercial Windows programs available which include a stellar database, telescope control for  
computerized telescopes like the Meade LX200, and CCD camera functions in an integrated  
package.  
6.7.2. Image Processing Software  
There are a host of image processing software packages capable of reading and processing TIFF  
files and new packages are being developed which will read and process SBIG image formats  
as well. In addition to commercial software, on-line services like CompuServe and GEnie have  
many public domain and shareware programs available through their Astronomy interest  
groups.  
6.7.3. Getting Hardcopy  
One older way of producing a hard copy of images taken with the CCD camera is to simply  
take a photograph of the computer screen while the image is displayed. It is best to use a long  
lens and step back from the monitor to reduce the appearance of the curvature of the edges of  
Page 41  
Section 6 - Accessories for your CCD Camera  
the screen, and use an exposure longer than 1/30th of a second to avoid the video refresh rate  
of your monitor. Darken the room, and use a brighter background than is visually optimum.  
For the best quality hard copy, save the files in TIFF format and send a copy of the file on a  
disk to a photo lab which offers printing of digital images. The Windows version of CCDOPS  
allows for printing of the images, and there are a number of third party software programs for  
the PC such as Pizzaz Plus which will capture and print the display on your computer screen.  
These programs, however, do not produce very detailed prints and are useful only to a very  
limited degree. Recently, photo quality color desktop printers have become commonplace.  
Many of these printers will do a reasonable job with commercial image processing programs  
such as Adobe Photoshop.  
6.8. SBIG Technical Support  
If you have any unanswered questions about the operation of your CCD camera system or  
have suggestions on how to improve it please don't fail to contact us. We appreciate all your  
comments and suggestions. Additionally if you are interested in writing software supporting  
SBIG cameras, we offer technical support regarding our file formats found in Appendix B, and  
Technical Notes regarding the camera command protocol which we will make available upon  
request.  
Page 42  
Section 8 - Glossary  
7.  
Common Problems  
This section discusses some of the more common problems others have encountered while  
using our CCD cameras. You should check here if you experience difficulties, and if your  
problem still persists please contact us to see if we can work it out together.  
Achieving Good Focus - Achieving a good focus is one of the most difficult areas in working  
with CCD cameras due to the lack of real time feedback when focusing. Focus can take  
a good deal of time, and as with all forms of imaging, focus is critical to getting the most  
out of your camera.  
Once you have achieved a good focus with your system, it can be very useful for  
future observing sessions to scribe an eyepiece or mark down or log positions of each  
component so the next time you will at least be close to focus at the start.  
If you know where the focal plane lies for your telescope, you can use Table 3.1  
to calculate exactly where the CCD is with respect to your system. By placing the CCD  
close to the focal plane initially, you can save a lot of time.  
The best kind of object to focus on is a star. As you converge towards focus,  
more light from the star will be concentrated onto one pixel. Thus, watching your peak  
reading while focusing and focusing for a maximum reading is a good way to get best  
focus. This is how we do it. It helps to have a dial or indicator on the focus knob so  
you can rapidly return to the best point after going through focus.  
Elongated Guided Images - When using Track and Accumulate or Self Guiding, if you notice  
guiding errors resulting in elongated star images, you are probably using too long a  
snapshot time. If the snapshot time is longer than the amount of time your drive can  
track unguided with acceptable guiding errors, you will see elongated stars in your  
final images. If your snapshot times are getting down to 30 seconds or less you should  
improve your drive.  
If you are using your camera as an autoguider for film photography and are noticing  
unacceptable guiding errors, please check the following before calling SBIG:  
1. Can you move the telescope using the Move command? This is an  
indicator as to whether or not you are properly connected to your  
drive system via the relay cable from the CPU.  
2. Be sure that your calibration time gives at least 10 to 50 pixels of  
movement for each step of the Calibrate Track command.  
3. Check for flexure between the CCD camera head and your system.  
Check for flexure between the guide scope or off-axis guider and  
your telescope system. This is a very common source of guiding  
errors. A very small movement of the CCD head with respect to the  
guide scope during an exposure can cause unacceptable streaking.  
4. If your mount is stable, try longer exposure times while tracking to  
average out the atmospheric effects.  
Finding Objects - The size of the CCD used in the ST-7 (roughly 8mm diagonal) can make  
finding objects a little difficult. If you experience this problem you might try the  
following suggestions.  
Page 43  
Section 8 - Glossary  
The easiest method of finding objects is to use a reticule eyepiece, if the object is  
bright enough to see. Pull the CCD optical head from the eyepiece holder and insert a  
12-20mm eyepiece, focussing the eyepiece by sliding it in and out of the eyepiece  
holder, not by adjusting the telescope's focus mechanism. Center the object carefully (to  
within 10% of the total field) and then replace the CCD optical head. Since the head  
was fully seated against the eyepiece holder when you started, fully seating it upon  
replacement will assure the same focus.  
If the object is too dim to see visually you will have to rely on your setting  
circles. Go to a nearby star or object that is easily visible and center that object in the  
CCD image. Calibrate your RA setting circle on the known object's RA and note any  
DEC errors. Reposition the telescope at the intended object, using the correct RA  
setting and the same DEC offset noted with the calibration object. Try a ten second or  
one minute exposure and hopefully you will have winged the object. If not you will  
have to hunt around for the object. You can use the Focus mode in Low resolution  
mode for this and hopefully you won't have to search too far. Check in DEC first, as  
DEC setting circles are often smaller and less accurate.  
Telescope Port doesn't Move Telescope - If you find the camera is not moving the telescope  
for Tracking, Track and Accumulate or Self Guiding you should use the Move  
Telescope command with a several second period to isolate the problem down to a  
specific direction or directions. If you set the Camera Resolution to the Low mode in  
the Camera Setup Command, you can move the telescope and Grab an image fairly  
quickly to detect movement of the telescope pointed at a moderately bright star. Try  
each of the four directions and see which ones move and which ones don't. At this  
point the most likely culprit is the hand controller modification. Trace the signals from  
the camera's telescope connector back through the hand controller, paying particular  
attention to the offending wires.  
Can't Reach Low Setpoint Temperatures - If you find that the camera isn't getting as cold as  
expected the problem is probably increased ambient temperatures. While these  
cameras have temperature regulation, they still can only cool a fixed amount below the  
ambient temperature (30 to 40 °C). Lowering the ambient temperature allows the  
cameras to achieve lower setpoint temperatures.  
CCD Frosts - If your camera starts to frost after a year of use it's time to regenerate the  
desiccant as described in Appendix C. This is a simple matter of unscrewing the  
desiccant container and baking it (without the little O-ring) in an oven at 350°F for 4  
hours.  
No Image is Displayed - Try the Auto Contrast setting or use the crosshairs to examine the  
image pixel values and pick appropriate values for the Background and Range  
parameters.  
Horizontal Faint Light Streaks in Image - some PCs apparently have the mouse generate non-  
maskable interrupts when moved. These interrupts can slightly brighten the line being  
read out. If this occurs, do not move the mouse during read out.  
A/D Timeout or Other Errors - Some compters require that you configure the parallel port for  
"Standard" mode instead of "ECP" or "EPP" mode. These two enhanced modes can  
cause A/D Timeout and other parallel errors. Refer to you computer manual for how  
to use the BIOS Setup utility to configure your parallel port.  
Page 44  
Section 8 - Glossary  
8.  
Glossary  
Antiblooming Gate - When a CCD pixel has reached its full well capacity, electrons can  
effectively spill over into an adjoining pixel. This is referred to as blooming. Kodak  
CCDs with the antiblooming option can be used to help stop or at least reduce  
blooming when the brighter parts of the image saturate.  
Astrometry - Astrometry is the study of stellar positions with respect to a given coordinate  
system.  
Autoguider - All SBIG CCD cameras have auto guiding or "Star Tracker" functions. This is  
accomplished by using the telescope drive motors to force a guide star to stay precisely  
centered on a single pixel of the CCD array. The camera has four relays to control the  
drive corrector system of the telescope. The CCD camera head is installed at the guide  
scope or off axis guider in place of a guiding eyepiece.  
CCD - The CCD (Charged Coupled Device) is a flat, two dimensional array of very small light  
detectors referred to as pixels. Each pixel acts like a bucket for electrons. The electrons  
are created by photons (light) absorbed in the pixel. During an exposure, each pixel fills  
up with electrons in proportion to the amount of light entering the pixel. After the  
exposure is complete, the electron charge buildup in each pixel is measured. When a  
pixel is displayed at the computer screen, its displayed brightness is proportional to the  
number of electrons that had accumulated in the pixel during the exposure.  
Dark Frame - The user will need to routinely create image files called Dark Frames. A Dark  
Frame is an image taken completely in the dark. The shutter covers the CCD. Dark  
Frames are subtracted from normal exposures (light frames) to eliminate fixed pattern  
and dark current noise from the image. Dark Frames must be of the same integration  
time and temperature as the light frame being processed.  
Dark Noise - Dark Noise or Dark Current is the result of thermally generated electrons  
building up in the CCD pixels during an exposure. The number of electrons due to  
Dark Noise is related to just two parameters; integration time and temperature of the  
CCD. The longer the integration time, the greater the dark current buildup. Conversely,  
the lower the operating temperature, the lower the dark current. This is why the CCD is  
cooled for long integration times. Dark noise is a mostly repeatable noise source,  
therefore it can be subtracted from the image by taking a "Dark Frame" exposure and  
subtracting it from the light image. This can usually be done with very little loss of  
dynamic range.  
Double Correlated Sampling - Double Correlated Sampling (DCS) is employed to lower the  
digitization errors due to residual charge in the readout capacitors. This results in lower  
readout noise.  
False Color - False Color images are images that have had colors assigned to different  
intensities instead of gray levels.  
FITS Image File Format - The FITS image file format (which stands for Flexible Image  
Transport System) is a common format supported by professional astronomical image  
processing programs such as IRAF and PC Vista. CCDOPS can save image files in this  
format but can not read them..  
Page 45  
Section 8 - Glossary  
Flat Field - A Flat Field is a image with a uniform distribution of light entering the telescope.  
An image taken this way is called a flat field image and is used with CCDOPS to correct  
images for vignetting.  
Focal Reducer - A Focal Reducer reduces the effective focal length of an optical system. It  
consists of a lens mounted in a cell and is usually placed in front of an eyepiece or  
camera. With the relatively small size of CCDs compared to film, focal reducers are  
often used in CCD imaging.  
Frame Transfer CCDs - Frame Transfer CCDs are CCDs that have a metal mask over some  
portion (usually half) of the pixel array. The unmasked portion is used to collect the  
image. After the exposure is complete, the CCD can very quickly shift the image from  
the unmasked portion of the CCD to the masked portion, thus protecting the image  
from light which may still be impinging on the CCD. This acts as an electronic shutter.  
Full Well Capacity - Full Well Capacity refers to the maximum number of electrons a CCD  
pixel can hold. This number is usually directly proportional to the area of the pixel.  
Histogram - The Histogram is a table of the number of pixels having a given intensity for each  
of the possible pixel locations of the image file. Remember that, in the end, the image  
file is nothing more than a list of pixel values, one for each CCD pixel. These value  
numbers can be displayed in two formats; as a table or plotted as a graph.  
Light Frame - The Light Frame is the image of an object before a Dark Frame has been  
subtracted.  
Path/Filter - For PC users the Path setting tells CCDOPS which directory and disk drive to find  
and store the image files. In essence, it defines the current directory. The filter controls  
which image files are shown in the Open directory. Only files that match the filter are  
shown. For example, a filter setting of "*.ST6" will only show files that have an "ST6"  
extension.  
Photometry - Photometry is the study of stellar magnitudes at a given wavelength or  
bandpass.  
Pixel Size - The smallest resolution element of a CCD camera is the CCD pixel. The pixel sizes  
for each of the SBIG cameras are as follows:  
Camera  
Tracking CCD13.75 x 16  
ST-5C 10 x 10  
ST-237/237A 7.4 x 7.4  
Pixel Size (microns)  
STV  
ST-6  
14.8 x 14.8  
23 x 277  
ST-7E/ST-8E 9 x 9  
ST-9E  
ST-10E  
ST-1001E  
20 x 20  
6.8 x 6.8  
24 x 24  
7
The TC241 CCD used in the ST-6 actually has 11.5 x 27 micron pixels. The CPU combines two 11.5  
micron pixels to make the 23 micron pixel . This gives pixel with a more square aspect ratio.  
Page 46  
Section 8 - Glossary  
Planet Mode - Planet Mode is the most useful way to achieve focus. When you select Planet  
mode, a full frame is exposed, downloaded, and displayed on the computer monitor. A  
small window can be placed anywhere in the image area and the size of the window  
can be changed. Subsequent downloads will be of the area inside the box resulting in a  
much faster update rate.  
Quantum Efficiency - Quantum Efficiency refers to the fractional number of electrons formed  
in the CCD pixel for a given number of photons. Quantum Efficiency is usually plotted  
as a function of wavelength.  
Readout Noise - Readout noise is associated with errors generated by the actual interrogation  
and readout of each of the CCD pixels at the end of an exposure. This is the result of  
fixed pattern noise in the CCD, residual charge in the readout capacitors and to a small  
extent the noise from the A/D converter and preamplifier.  
Resolution Mode - The resolution of a CCD camera is determined by pixel size. Pixel size can  
be increased by combining or binning more than one pixel and displaying it as one  
pixel. Doing so decreases the effective resolution but speeds up the download time of  
the image. Maximum resolution is determined by the size of the individual CCD pixel.  
The ST-7E/8E/9E can run in High, Medium, Low and Auto resolution modes.  
Response Factor - Response Factor is a multiplier used by CCDOPS to calibrate CCDOPS to a  
given telescope for photometric calculations. The Response Factor multiplied by 6700 is  
the number of photoelectrons generated in the CCD for a 0th magnitude star per second  
per square inch of aperture.  
Saturation - Saturation refers to the full well capacity of a CCD pixel as well as the maximum  
counts available in the A/D converter. The pixel is saturated when the number of  
electrons accumulated in the pixel reaches its full well capacity. The A/D is saturated  
when the input voltage exceeds the maximum. The saturation values for the various  
cameras are shown in the table below.  
Full Well  
Capacity  
Maximum  
A/D  
Counts  
Electrons  
per A/D  
Camera  
8
count  
ST-5C  
ST-237  
ST-237A  
ST-6  
50,000 e-  
20,000 e-  
20,000 e-  
400,000 e-  
50,000 e-  
100,000 e-  
150,000 e-  
180,000 e-  
77,000 e-  
180,000e-  
65,535  
4,095  
1.25  
3.4  
0.7  
6.7  
2.3  
2.3  
1.3  
2.8  
1.5  
2.8  
65,535  
65,535  
20,000  
65,535  
65,535  
65,535  
65,535  
65,535  
9
ST-7E/ST-8E ABG High Res  
ST-7E/ST-8E Med/Low Res  
ST-7E/8E/9E/10E Tracking CCD  
ST-9E  
ST-10E  
ST-1001E  
8
9
-
The e /Count shown are for the High Resolution readout mode. These numbers can be different for other  
readout modes. Use the Parameters command in the Display menu to see the actual value for images acquired  
in other readout modes.  
The non-ABG versions of the ST-7E/8E have twice the full well capacity of the values shown. ABG version  
may also vary from camera to camera with maximum counts being as high as ~30,000.  
Page 47  
Section 8 - Glossary  
Sky Background - The sky background illumination or brightness is the number of counts in  
the image in areas free of stars or nebulosity and is due to city lights and sky glow.  
High levels of sky background can increase the noise in images just like dark current.  
For some objects deep sky filters can be used to reduce the sky background level.  
Seeing - Seeing refers to the steadiness and the clarity of the atmosphere during an observing  
session.  
TE Cooler - The TE Cooler is a Thermal Electric cooling device used to cool the CCD down to  
operating temperature. The CCD is mounted to the TE Cooler which is mounted to a  
heat sink, usually the camera head housing.  
TIFF Image File Format - The TIFF image file format (which stands for Tagged Interchange  
File Format) was developed jointly by Microsoft and Aldus Corporations to allow easy  
interchange of graphics images between programs in areas such as presentation and  
desktop publishing. CCDOPS can save image files in this format but it can not read  
them.  
Track and Accumulate - The Track and Accumulate function is a SBIG patented feature of  
CCDOPS that allows the user to automatically co-register and co-add (including dark  
frame subtraction) a series of images of an object. These exposures can be taken  
unguided as long as the "Snapshot time" does not exceed the length of time before  
tracking errors of your clock drive become apparent. This allows you to image and track  
without guiding or the need to connect the CCD Relay port to your drive motors.  
Track List - The Track List is an ASCII file generated by CCDOPS during a Track and  
Accumulate session. The Track List logs all the corrections made by CCDOPS for each  
of the images. Track lists are required when flat fielding Track and Accumulate images.  
Tri-Color - Tri-Color refers to color images created using three different colors mixed into a  
balanced color image using red, green and blue filters. An object is imaged three times,  
once with each color filter. The three images are then co-added and color balanced with  
the appropriate software.  
Vignetting - Vignetting is obstruction of the light paths by parts of the instrument. It results in  
an uneven illumination of the image plane. The effects of vignetting can be corrected  
using flat field images.  
Page 48  
Appendix A - Connector Pinouts  
A.  
Appendix A - Connector ad Cables  
This appendix describes the various connectors and cables used with the ST-  
7E/8E/9E/10E/1001E.  
A.1. Appendix A - Connector Pinouts  
Tables A1 and A2 below show the pin-outs of the Telescope and Power connectors on the ST-  
7E/8E/9E.  
Pin Number  
Function  
1
2
3
4
Chassis Ground  
CFW-8 Pulse/AO-7 Data Out  
Plus X (Active Low Open Drain)9  
Plus Y (Active Low Open Drain)  
Signal Ground  
5
6
7
8
9
Minus X (Active Low Open Drain)  
Minus Y (Active LowOpen Drain)  
+12 Volts Out (100mA max)  
+5 Volts Out (300 mA max)  
Chassis Ground  
Shell  
Table A1 Telescope Connector  
Pin Number  
6, Shell  
5
4
3
2
1
Function  
Earth Ground  
DC Ground  
-12V DC, 100mA  
No contact  
+12V DC, 500mA  
+5V DC, 2A  
Table A2 Power Connector  
+12VDC on outside  
Ground in Center  
Mating plug is 5.5mm  
outside and 2.1mm inside  
Figure A1 - Cooling Booster DC Power Jack  
A.2. SBIG Tracking Interface Cable (TIC-78)  
Many of the newer telescopes have a phone-jack connector on the drive corrector for  
connecting directly to the ST-7E/8E/9E/10E/1001E Camera's Telescope Port. These include  
9
The Open Drain outputs can sink 100 mA maximum  
Page 49  
Appendix A - Connector Pinouts  
the Celestron Ultima, Losmandy CG11 and Meade LX-200. You can interface these telescopes  
to the Telescope port with our TIC-78 (Tracking Interface Cable), or you can make your own  
cable. Figure A1 below shows the pinouts on these telescopes.  
Hand Controller  
Button  
Modular Phone  
Wire Color  
Telescope Port Pin  
Special  
Common  
Left  
White  
Black  
Red  
Green  
Yellow  
Blue  
None  
5
6
7
4
3
Down  
Up  
Right  
Figure A1 - CCD Connector for TIC Mating  
Page 50  
Appendix B - File Formats  
B.  
Appendix C - Maintenance  
This appendix describes the maintenance items you should know about with your CCD  
camera system.  
B.1. Cleaning the CCD and the Window  
The design of SBIG cameras allows for cleaning of the CCD. The optical heads are not  
evacuated and are quite easy to open and clean. When opening the CCD chamber, one should  
be very careful not to damage the structures contained inside.  
To open the CCD Chamber, remove the six screws that hold the 5 inch front cover in  
place. Remove the six screws and lift the front cover, exposing the structures inside. There is a  
rubber O-Ring that sets in the groove on the top of the Chamber housing.  
The CCD array is protected by a thin cover glass that can be cleaned with Q-Tips and  
Isopropyl Alcohol. Do not get alcohol on the shutter. Dust on the CCD should be blown off. Use  
alcohol only if necessary. The optical window of the chamber housing can be cleaned the same  
way. When reinstalling the chamber housing, be very careful to make sure the O-ring is in the  
groove when seated.  
B.2. Regenerating the Desiccant  
This section describes the regeneration procedure for the desiccant used in the  
ST-7E/8E/9E/10E/1001E. The desiccant absorbs moisture in the CCD chamber, lowering the  
dew point below the operating temperature of the cooled CCD, thus preventing the formation  
of frost. The desiccant is contained in a small cylindrical plug that screws into the chamber  
from the rear. In normal operation the useful life of the desiccant is over a year. If the CCD  
chamber is opened often, the desiccant should be regenerated when frosting is noticed. Follow  
the procedure below to regenerate the desiccant:  
1. Unscrew the brass desiccant container from the rear of the camera and  
remove the O-ring.  
2.. Plug the resulting hole by screwing in the supplied bolt or plug 2 or 3 turns.  
Finger tight is adequate. Don't put a wrench on it.  
2. Heat the dessicant container in an oven at 350°F (175 deg C) for 4 hours.  
The solder used to seal the can melts at 460 degrees F, so be sure to stay at  
least 50 degrees below this number. Preheating the oven to avoid hot spots  
is advised.  
3. Replace the desiccant container into the rear of the camera, being careful to  
reinstall the O-ring and insure that it does not get pinched.  
4. Expect the camera to take an hour or two to reach the frost free state. If it  
does seem to frost and you need to capture images, reduce your cooling to  
the zero degree C range - the CCD dark current will still be quite low.  
Page 51  
Appendix C - Capturing a Good Flat Field  
C.  
Appendix C - Capturing a Good Flat Field  
This appendix describes how to take a good flat field. A good flat field is essential for  
displaying features little brighter than the sky background. The flat field corrects for  
pixel non-uniformity, vignetting, dust spots (affectionately called dust doughnuts), and  
stray light variations. If the flat field is not good it usually shows up as a variation in  
sky brightness from on side of the frame to the other.  
C.1. Technique  
The first consideration in capturing a flat field is to use the telescope-CCD combination  
in exactly the configuration used to collect the image. This means you probably have to  
capture the flat field at the telescope. Do not rotate the head between image and flat  
field, since the vignetting is usually slightly off center. Do not be tempted to build a  
little LED into the telescope or camera for doing flat fields; it doesn't work at all. The  
dust debris shadows would be different!  
Arrange a light source such as a flashlight, two white cards, the telescope and  
CCD as shown in Figure D-1.  
Figure D-1: Flat Field Geometry  
Telescope  
CCD  
Flat White  
Surface  
Flashlight  
Flat White  
Surface  
The key aspects of this geometry are that the reflection off two diffuse surfaces is used,  
and the large flat surface is square to the illumination from the small flat surface. When  
we do this, the first flat surface is typically a white T-shirt worn by the operator! Take  
care that no apparent shadows are cast onto the larger flat white surface. Use an  
exposure at the camera that yields an average light level equal to about half of full scale.  
Page 53  
Index  
A/D converter, 8, 27  
accessories, 41  
adaptive optics, 40  
antiblooming, 29, 45  
Antiblooming Gate (def), 45  
AO-7, 40  
PC Setup, 16  
communications link, 16  
connector  
Telescope, 49  
connector (CPU)  
RELAYS, 23, 37  
astrometric measurements, 45  
Astrometry (def), 45  
astrophotography, 12  
atmospheric effects, 17  
auto contrast, 18  
Auto Grab Command, 21  
autoguider, 8, 20, 23, 36, 45  
Autoguider (def), 45  
background parameter, 18  
battery operation, 15, 23  
binning, 10  
cooling, 8, 48  
co-register images, 20, 23  
crop, 18  
crosshairs, 18  
dark current, 8  
dark frame, 8, 9, 15, 18  
Dark Frame (def), 45  
Dark Noise (def), 45  
Dark Subtract Command, 19  
dark subtracted, 9, 15, 18  
DCS, 9, 45  
Calibrate Track Command, 21, 37  
calibration star, 37  
camera lens adapter, 33, 40  
Camera Setup Commands, 20  
CCD, 7, 11  
Declination, 15, 23, 37, 44  
desiccant, 51  
Diffuse Magnitude, 19  
diffuser, 16  
Dim mode, 19  
(def), 45  
cameras, 5, 6, 8, 15, 23, 26, 27, 28  
cleaning, 51, 53  
detector, 5, 7, 23  
orientation, 15  
displaying images, 18  
double correlated sampling, 9  
Double Correlated Sampling (def), 45  
drive corrector, 20  
electromechanical shutter, 27  
electromechanical vane, 27  
electronic images, 12  
electronic shutter, 27  
Establish COM Link Command, 16  
eyepiece, 17, 18  
CCDCOLOR software, 21  
CCDOPS software, 8  
centering objects, 18  
cleaning the CCD, 51, 53  
clock drivers, 8  
co-add images, 5, 20, 23  
cold pixels, 19  
color filter wheel, 21, 40  
color images, 13, 21  
color table, 45  
eyepiece holder, 15  
eyepiece projection, 40  
eyepiece tube, 17  
False Color (def), 45  
field of view, 11, 28, 29, 37  
film, 5, 10, 11  
commands  
Auto Grab, 21  
Calibrate Track, 21, 37  
Dark Subtract, 19  
Focus, 17, 19  
Grab, 18, 33  
film grain, 10  
filter wheel, 13  
filters, 13, 19, 21, 40  
finding objects, 18  
fine focus, 17  
Track, 37  
FITS format (def), 45  
flat field, 10, 19, 34, 35  
Flat Field (def), 46  
flip, 19  
Track and Accumulate, 20, 33, 37  
Tracking Parameters, 37  
Commands  
Establish COM Link, 16  
focal length, 30  
55  
Index  
focal reducer, 40  
Focal Reducer (def), 46  
focus  
optical head, 15, 23  
optical head orientation, 16  
O-ring, 8  
path (def), 46  
path/filter (def), 46  
PC, 8  
PC Setup Command, 16  
PEC drive correctors, 20, 37  
photometric measurements, 46  
Photometry (def), 46  
pixel nonuniformity, 19  
pixel size, 10, 28  
Dim mode, 18, 19  
fine, 17  
Full frame mode, 17  
Full Frame mode, 19  
peak, 17  
Planet mode, 17, 19, 46  
Focus Command, 17, 19  
focus mode, 17, 19  
Focusing, 16  
Frame Transfer CCDs, 6  
Frame Transfer CCDs (def), 46  
frost, 8  
Full Frame CCDs, 6  
full well capacity, 29  
Full Well Capacity (def), 46  
Grab Command, 18, 33  
graphics, 18  
Pixel Size (def), 46  
pixel uniformity, 10  
planet mode, 19  
Planet mode (def), 46  
planets, 5, 16, 30, 33  
POWER connector, 15  
power supply, 8, 15  
power supplyr, 23  
PPEC drive correctors, 20, 37  
preamp, 8, 23  
printing images, 12, 41  
push to make switch modification, 25  
push to make/break switch modification,  
25  
Quantum Efficiency (def), 47  
range parameter, 18  
readout  
guiding, 11  
guiding error, 20, 37  
hand controller, 18, 20, 23, 24  
hermetic chamber, 8  
Histogram (def), 46  
host computer, 8, 15, 23  
hot pixels, 19  
IBM PC, 8  
image processing, 8, 12, 18, 41  
interference, ii  
electronics, 23  
IR blocking filter, 22  
joystick, 25  
joystick modification, 26  
library of dark frames, 18, 34  
light frame, 9, 15, 18  
light frame (def), 46  
Link, 16  
noise, 29  
Noise (def), 47  
register, 6  
regenerating desiccant, 51  
RELAYS connector, 23, 37  
resolution, 10, 20, 34  
Resolution (def), 47  
Response factor (def), 47  
Right Ascension, 15, 23, 37, 44  
Saturation (def), 47  
scribing the eyepiece, 17  
seeing, 10, 17, 20, 30  
Seeing (def), 48  
LPT port, 15, 16, 23  
magnitude  
diffuse, 19  
stellar, 19  
maintenance, 51, 53  
microcontroller, 8  
modifying hand controller, 24  
moon, 16, 30, 33  
self guiding, 23  
separations, 19  
negative image, 18  
nosepiece, 15  
observatory, 23  
setup, 15  
SGS-Self Guided Spectrograph, 41  
sharpen, 19  
56  
Index  
shutter, 8  
signal to noise ratio, 22, 29  
sky background, 5, 20  
smoothing, 19  
snapshot, 23  
software, 26, 41  
spectral range, 5  
spectrograph, 41  
Status Window, 16  
Link field, 16  
stellar magnitude, 19  
stellar temperature, 19  
super pixel, 19  
taking images, 18  
TE cooler, 8  
TE Cooler, 7  
TE Cooler (def), 48  
Technical Support, 42  
telescope, 10, 15, 18, 20  
Telescope connector, 49  
telescope hand controller, 18, 20, 23, 24  
temperature regulation, 9, 18, 27  
thermistor, 9  
TIC, 49  
TIC-78, 24  
TIFF format (def), 48  
Track and Accumulate (def), 48  
Track and Accumulate Command, 20, 33,  
37  
Track and Accumulate mode, 23, 35  
Track Command, 37  
track list, 35  
Track List (def), 48  
track mode, 23  
tracking interface cable (TIC), 24, 49  
Tracking Parameters Command, 37  
Tri-Color (def), 48  
tri-color images, 40  
T-Thread, 40  
vane, 27  
vignetting, 10, 19  
Vignetting (def), 48  
Windows, 41  
zoom, 18  
57  

Rockford Fosgate FRC3204T User Manual
Radio Shack 23 154 User Manual
PYLE Audio PLD182 User Manual
Philips GOGEAR SA4310 User Manual
Olympus 227580 User Manual
MB QUART MBQ12SB4 User Manual
Leica SBC RU 430 User Manual
Kodak Digital Frame 1 User Manual
Kicker KS690 User Manual
JVC KD LH3100 User Manual